FDA Adverse Event Malfunction Summary report: N

TAXUS EXPRESS2 PACLITAXEL-ELUTING CORONARY STENT SYSTEM

MDR report key: 1170151 · Received September 11, 2008

Report

Report Number
2134265-2008-02603
Event Type
Malfunction
Date Received
September 11, 2008
Date of Event
July 15, 2008
Report Date
August 19, 2008
Manufacturer
BOSTON SCIENTIFIC CORPORATION
Product Code
NIQ
PMA / PMN Number
P030025
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

FOLLOWING A DETAILED DEVICE ANALYSIS, IT WAS NOTED THAT THE CONDITION OF THE RETURNED DEVICE WAS CONSISTENT WITH THE COMPLAINT INCIDENT. THE DEVICE WAS RETURNED FOR ANALYSIS AND INITIAL VISUAL EXAMINATION OF THE RETURNED DEVICE REVEALED PROXIMAL STENT DAMAGE. TWO OF THE STRUTS WERE BENT BACK DISTALLY. THIS TYPE OF DAMAGE IS CONSISTENT WITH THE DEVICE ENCOUNTERING SOME FORM OF RESTRICTION ON THE WITHDRAWAL OF THE DEVICE. NO KINKS OR DAMAGE WERE FOUND ON THE HYPOTUBE. THE BALLOON AND TIP SECTIONS OF THE DEVICE WERE VISUALLY AND MICROSCOPICALLY EXAMINED, AND NO ISSUES WERE NOTED WITH THEIR PROFILES THAT COULD HAVE POTENTIALLY CONTRIBUTED TO THE COMPLAINT INCIDENT. THE BALLOON WAS TIGHTLY WRAPPED AND WAS NOT SUBJECTED TO ANY POSITIVE PRESSURE. A 0.015 INCH PRODUCT MANDREL WAS INSERTED THROUGH THE LUMEN WITH NO RESTRICTIONS NOTED. ALL UNITS SHIPPED FOR THE BATCH CONFORMED TO PRODUCT SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE OF THIS EVENT IS OPERATIONAL CONTEXT.

Description of Event or Problem · 1

THIS COMPLAINT IS NOW REPORTABLE BASED ON THE ANALYSIS COMPLETED ON 08/19/2008. IT WAS REPORTED THAT DURING A CORONARY DRUG ELUTING STENTING TREATMENT PROCEDURE, THE STENT COULD NOT CROSS THE 90% STENOSED, SEVERELY CALCIFIED, AND EXTREMELY TORTUOUS DISTAL CIRCUMFLEX LESION. NO PATIENT INJURIES OR COMPLICATIONS WERE REPORTED. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. HOWEVER, THE RETURNED PRODUCT CONFIRMED THAT THERE WAS STENT DAMAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TAXUS EXPRESS2 PACLITAXEL-ELUTING CORONARY STENT SYSTEM NIQ - STENT, CORONARY, DRUG-ELUTING NIQ BOSTON SCIENTIFIC CORPORATION 2.75X32MM 11623782

Patients

Seq Age Sex Outcome Treatment
1