FDA Adverse Event Malfunction Summary report: N

DXH500 HEMATOLOGY INSTRUMENT

MDR report key: 11700487 · Received April 21, 2021

Report

Report Number
1061932-2021-00060
Event Type
Malfunction
Date Received
April 21, 2021
Date of Event
March 15, 2021
Report Date
April 21, 2021
Manufacturer
BECKMAN COULTER IRELAND INC.
Product Code
GKZ
UDI-DI
15099590671242
PMA / PMN Number
K181475
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FIELDS SERVICE ENGINEER (FSE) COULD NOT IDENTIFY A SPECIFIC MALFUNCTION OR FAULTY PART ASSOCIATED TO THIS EVENT. THERE WAS NO REPAIR OR REPLACEMENT OF A COMPONENT PERFORMED. THE CUSTOMER DID NOT CALL BACK WITH A RECURRENCE OF THE EVENT AS OF (B)(6)2021. SECTION A PATIENT INFORMATION WAS NOT PROVIDED BY THE CUSTOMER. BEC INTERNAL IDENTIFIER CASE: (B)(6).

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THEIR DXH 500 HEMATOLOGY INSTRUMENT GENERATED ERRONEOUSLY HIGH PLATELET (PLT) RESULTS ON ONE PATIENT WITH NO FLAGGING OR ALERT ON PLT CLUMPS THE PATIENT SAMPLE WAS INITIALLY RUN ON A DXH 800 HEMATOLOGY INSTRUMENT WHICH GENERATED R FLAGS AND A PLT RESULT OF 14.0 AND PROMPTED THE OPERATOR TO REVIEW THE RESULTS. THEY RERAN THE PATIENT ON THE DXH 500 GENERATING MUCH HIGHER PLT RESULTS (125.4) PROMPTING FOR A MANUAL SLIDE REVIEW THAT COINCIDED WITH THE INITIAL RUN ON THE DXH 800 AND CONFIRMED PLT CLUMPS ON THE SAMPLE. THERE WAS NO INDICATION OF ERRONEOUS RESULTS BEING REPORTED OUTSIDE THE LABORATORY RAW DATA FROM THE INSTRUMENT WAS REQUESTED BUT NOT PROVIDED. PATIENT RESULTS WERE PROVIDED AND ASSESSED FOR THIS EVENT. THERE WAS NO REPORT OF CHANGE OR EFFECT TO PATIENT TREATMENT. THERE WAS NO ADVERSE EVENT REPORTED AS A RESULTS OF THIS OCCURRENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
597542 DXH500 HEMATOLOGY INSTRUMENT COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER IRELAND INC. TN,DXH 500 15099590671242

Patients

Seq Age Sex Outcome Treatment
1