FDA Adverse Event Malfunction Summary report: N

QUIDEL LYRA DIRECT SARS-COV-2 ASSAY (M124)

MDR report key: 11698416 · Received April 20, 2021

Report

Report Number
1000122536-2021-00010
Event Type
Malfunction
Date Received
April 20, 2021
Date of Event
February 9, 2021
Report Date
April 20, 2021
Manufacturer
QUIDEL CORPORATION
Product Code
QJR
UDI-DI
30014613339144
PMA / PMN Number
EUA200423
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATIONAL SUMMARY: REVIEW OF CUSTOMER DATA SHOWS THAT SAMPLE HI-02 YIELDED AN ATYPICAL CURVE IN ONE RUN WHICH GENERATED A POSITIVE RESULT DUE TO THE CURVE CROSSING THE SARS-COV-2 THRESHOLD. A SECOND RUN OF THE SAME SAMPLE SHOWS A FLAT CURVE IN THE SARS-COV-2 CHANNEL. GIVEN THE ATYPICAL AMPLIFICATION BEHAVIOR OF THE CURVE IN THE FIRST RUN, IT IS POSSIBLE THAT THERE COULD BE SOME INTERFERENCE COMING FROM THE SAMPLE. RETAIN TESTING NOT PERFORMED SINCE LOT # NOT PROVIDED BY CUSTOMER. WHILE QC WAS UNABLE TO DETERMINE THE ROOT CAUSE, THE COMPLAINT HAS BEEN DOCUMENTED AND WILL CONTINUE TO BE MONITORED.

Description of Event or Problem · 0

FALSE POSITIVE: CUSTOMER REPORTED POTENTIAL FALSE POSITIVE RESULT ON LYRA DIRECT SARS-COV-2.

Additional Manufacturer Narrative · 1

INVESTIGATIONAL SUMMARY: REVIEW OF CUSTOMER DATA SHOWS THAT SAMPLE HI-02 YIELDED AN ATYPICAL CURVE IN ONE RUN WHICH GENERATED A POSITIVE RESULT DUE TO THE CURVE CROSSING THE SARS-COV-2 THRESHOLD. A SECOND RUN OF THE SAME SAMPLE SHOWS A FLAT CURVE IN THE SARS-COV-2 CHANNEL. GIVEN THE ATYPICAL AMPLIFICATION BEHAVIOR OF THE CURVE IN THE FIRST RUN, IT IS POSSIBLE THAT THERE COULD BE SOME INTERFERENCE COMING FROM THE SAMPLE. RETAIN TESTING NOT PERFORMED SINCE LOT # NOT PROVIDED BY CUSTOMER. WHILE QC WAS UNABLE TO DETERMINE THE ROOT CAUSE, THE COMPLAINT HAS BEEN DOCUMENTED AND WILL CONTINUE TO BE MONITORED.

Description of Event or Problem · 1

FALSE NEGATIVE: CUSTOMER REPORTED POTENTIAL FALSE POSITIVE RESULT ON LYRA DIRECT SARS-COV-2.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
589607 QUIDEL LYRA DIRECT SARS-COV-2 ASSAY (M124) LYRA DIRECT SARS-COV-2 ASSAY QJR QUIDEL CORPORATION CORONAVIRUS 30014613339144

Patients

Seq Age Sex Outcome Treatment
1