VACUETTE TUBE 3 ML 9NC COAGULATION SODIUM CITRATE 3,2% 13X75
Report
- Report Number
- 1125230-2021-00036
- Event Type
- Malfunction
- Date Received
- April 20, 2021
- Report Date
- April 20, 2021
- Manufacturer
- GREINER BIO-ONE NA INC.
- Product Code
- GIM
- PMA / PMN Number
- K971221
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
COMPLAINT CO21-2100-197 RECEIVED 1RK 454334/B20113CT FOR EVALUATION. A CHECK OF QUALITY, PRODUCTION, AND MAINTENANCE RECORDS REVEALED NO DEVIATIONS IN RELATION TO THE REPORTED ERROR. SAMPLES WERE TESTED, ACCORDING TO GBO STANDARD TESTING PROCEDURES, WITH REGARDS TO CORRECT ASSEMBLY, LEVEL MARK, FILLING LEVEL AND DRAW VOLUME ACCORDING TO ISO 6710 'SINGLE USE CONTAINERS FOR VENOUS BLOOD SPECIMEN COLLECTION' AND CLSI GP39-A6 REGULATIONS FOR 'EVACUATED TUBES AND ADDITIVES FOR BLOOD SPECIMEN COLLECTION'. TUBES WERE VERIFIED TO BE CORRECTLY ASSEMBLED AND HAD THE CORRECT FILL GUIDELINE POSITION. GREINER FILL MARK PROVIDES A VISUAL CONTROL OPPORTUNITY FOR THE PHLEBOTOMIST AND FOR THE LAB PERSONNEL TO CHECK FOR PROPER VOLUME COLLECTION OF SPECIMEN. BOTH STANDARDS SPECIFY THE DRAW VOLUME TO BE WITHIN +/- 10% RANGE OF THE NOMINAL FILL VOLUME. NO VISUAL DEVIATION IN FILL VOLUME, SPORADIC LOW OR HIGH FILL, COULD BE OBSERVED IN THE TESTED SAMPLES. ALL TUBES TESTED FILLED WITHIN THE +/-10% TOLERANCE RANGE. THE COMPLAINT COULD NOT BE DUPLICATED.
CUSTOMER STATES TUBES ARE UNDERFILLING. CUSTOMER ALSO ADVICES THAT GLASS DISCARD TUBE IS DRAWN BEFORE COAG TUBE, TUBES ARE ONLY SUPPOSED TO BE DRAWN WITH STRAIGHT NEEDLE AND UNDERFILLED SPECIMENS COME PRIMARILY FROM THE ER (RN DRAWS). THE LOT NUMBER IS THE ONLY ONE THEY HAVE. UNKNOWN IF ISSUE HAS BEEN SEEN IN ANY OTHER LOT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 595978 | VACUETTE TUBE 3 ML 9NC COAGULATION SODIUM CITRATE 3,2% 13X75 | EVACUATED BLOOD COLLECTION TUBE | GIM | GREINER BIO-ONE NA INC. | 454334 | B20113CT |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|