FDA Adverse Event Malfunction Summary report: N

DRY SKIN PREP TRAY

MDR report key: 11696909 · Received April 20, 2021

Report

Report Number
3004932373-2021-00203
Event Type
Malfunction
Date Received
April 20, 2021
Date of Event
March 22, 2021
Report Date
October 13, 2021
Manufacturer
CAREFUSION, INC
Product Code
OJU
UDI-DI
10885403142192
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

ACCORDING TO THE SUPPLIER, LOT NUMBER WAS REVIEWED AND NO ISSUES WERE FOUND. A SAMPLE WAS PROVIDED BY THE CUSTOMER AND IT WAS OBSERVED THAT HAIR WAS ON THE COMPONENT . THE FAILURE MODE WAS CONFIRMED WITH THE SAMPLE RECEIVED. THE ROOT CAUSE IS DUE TO MANUFACTURING PERSONNEL AS DRESS CODE WAS NOT FOLLOWED AND CAUSED THE CONTAMINATION. AN EMPLOYEE AWARENESS TRAINING WAS PERFORMED AT THE PLANT TO ENSURE THIS IS AN ISOLATED INCIDENT. H3 OTHER TEXT : SEE NARRATIVE BELOW.

Description of Event or Problem · 0

MATERIAL NO.: 4461A, BATCH NO.: 0004173982. IT WAS REPORTED THAT A HAIR WAS FOUND IN THE TRAY BEFORE USE. PER EMAIL: CAN YOU FORWARD THIS QUALITY ISSUE TO THE BD REP. OR FOUND A HAIR IN THE DRY PREOP SKIN PREP TRAY. ECH # 007202, REF# 4461A, LOT # 004173982. IT WAS ATTACHED TO ONE OF THE SPONGE STICKS. SEE BELOW.

Additional Manufacturer Narrative · 1

(B)(4). (B)(6) INITIAL EMDR SUBMISSION. A FOLLOW UP EMDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

MATERIAL NO.: 4461A, BATCH NO.: 0004173982. IT WAS REPORTED THAT A HAIR WAS FOUND IN THE TRAY BEFORE USE. PER EMAIL: CAN YOU FORWARD THIS QUALITY ISSUE TO THE BD REP. OR FOUND A HAIR IN THE DRY PREOP SKIN PREP TRAY. ECH # 007202, REF# 4461A, LOT # 004173982. IT WAS ATTACHED TO ONE OF THE SPONGE STICKS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
592545 DRY SKIN PREP TRAY DRY SKIN PREP TRAY OJU CAREFUSION, INC 4461A 0004173982 10885403142192

Patients

Seq Age Sex Outcome Treatment
1 Other