TRACH FLEX DIC CUF 7.0MM
Report
- Report Number
- 9610530-2008-00041
- Event Type
- Malfunction
- Date Received
- September 18, 2008
- Date of Event
- August 18, 2008
- Report Date
- September 3, 2008
- Manufacturer
- SMITHS HEALTHCARE MFG SA DE C.V.
- Product Code
- BTO
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
SAME PT, WITH LOT 1212376, MANUFACTURED 10/2007. RESULTS EVAL - THE INVESTIGATION INTO THIS EVENT IS BASED ON HISTORY. NO SAMPLE WAS RETURNED DUE TO PT HAVING INFECTIOUS DISEASE, AND NO SAFE WAY TO SHIP SAMPLE BACK, OR TEST AFFECTED SAMPLES. REVIEW OF THE MFG RECORDS, FOR BOTH LOTS, REVEALED NO PROBLEMS DURING MANUFACTURE. A FURTHER REVIEW OF MFG RECORDS CONFIRMED THE PRESENCE OF AN INFLATION CHECK CARRIED OUT ON 100% OF THIS PRODUCT LINE PRIOR TO PACKAGING. PER REPORTER, THEY HAVE ONLY HAD ONE SIMILAR EVENT, WITH THIS CATALOG NUMBER, WITH A DIFFERENT PT. FIVE EVENTS WITH SAME PT, TWO LOTS. WITHOUT THE EVENT SAMPLES, THERE IS NO WAY TO DETERMINE POSITIVELY, HOWEVER, DUE TO HISTORY OF THIS TYPE OF REPORT, ROOT CAUSE LIKELY DUE TO PT ANATOMY. IT IS STATED THAT THE UNITS WERE TESTED PRIOR TO USE AND ONLY BECAME DEFLATED AFTER MORE THAN AN HOUR IN USE. AS SUCH, THIS INCIDENT IS UNLIKELY THE RESULT OF AN ASSEMBLY FAULT. WE HAVE DETERMINED THE ROOT CAUSE FOR THIS INCIDENT IS THAT THE CUFF BECAME SCRATCHED DURING INSERTION THROUGH THE STOMA AND SUBSEQUENTLY PUNCTURED FOLLOWING INSERTION. THOUGH THESE PRODUCTS ARE DESIGNED TO BE AS ROBUST AS FUNCTIONALLY POSSIBLE, PUNCTURE OF THE TUBE CUFF ON THE PT'S ANATOMY CAN BE EXPERIENCED.
USER ALLEGES ONE UNIT HAS PINHOLES IN ANTERIOR UPPER SIDE OF THE CUFF. TRACHEOSTOMY TUBE HAD BEEN IN PLACE IN EXCESS OF ONE HOUR. FIVE EVENTS WITH SAME PT. UNIT WAS PRETESTED PER THE INSTRUCTIONS FOR USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TRACH FLEX DIC CUF 7.0MM | 73 BTO - TRACHEOSTOMY TUBE, SINGLE-USE | BTO | SMITHS HEALTHCARE MFG SA DE C.V. | 1079100 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |