FDA Adverse Event
Malfunction
Summary report: N
HEATER-COOLER SYSTEM 3T
MDR report key: 11694065
·
Received April 20, 2021
Report
- Report Number
- 9611109-2021-00224
- Event Type
- Malfunction
- Date Received
- April 20, 2021
- Date of Event
- March 24, 2021
- Report Date
- April 20, 2021
- Manufacturer
- LIVANOVA DEUTSCHLAND
- Product Code
- DWC
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THERE WAS NO PATIENT INVOLVEMENT. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K191402). LIVANOVA (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T . THE INCIDENT OCCURRED IN (B)(6). A LIVANOVA FIELD SERVICE REPRESENTATIVE WAS DISPATCHED TO THE FACILITY TO INVESTIGATE THE DEVICE AND COULD CONFIRM THE REPORTED ISSUE. THE FAILURE WAS TRACED BACK TO A DEFECTIVE COOLING COMPRESSOR. THE DEVICE WILL NOT BE REPAIRED AND A REPLACEMENT UNIT HAS BEEN PROVIDED TO THE CUSTOMER.
Description of Event or Problem · 1
LIVANOVA (B)(4) RECEIVED A REPORT THAT A HEATER-COOLER SYSTEM 3T TRIPPED THE MAINS RCD DURING USE. THERE WAS NO REPORT OF PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 592053 | HEATER-COOLER SYSTEM 3T | CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS | DWC | LIVANOVA DEUTSCHLAND | 16-02-80 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |