VACUETTE TUBE 6 ML K2E K2EDTA 13X100 PINK CAP-PINK RING, NON-RIDGED
Report
- Report Number
- 1125230-2021-00035
- Event Type
- Malfunction
- Date Received
- April 20, 2021
- Report Date
- April 20, 2021
- Manufacturer
- GREINER BIO-ONE NA INC.
- Product Code
- JKA
- PMA / PMN Number
- K971236
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
COMPLAINT CO21-2100-194 RECEIVED 1RK 456279/B210233F FOR EVALUATION. A CHECK OF QUALITY, PRODUCTION, AND MAINTENANCE RECORDS REVEALED NO DEVIATIONS IN RELATION TO THE REPORTED ERROR. SAMPLES WERE TESTED, ACCORDING TO GBO STANDARD TESTING PROCEDURES, WITH REGARDS TO CORRECT ASSEMBLY, LEVEL MARK, FILLING LEVEL AND DRAW VOLUME ACCORDING TO ISO 6710 'SINGLE USE CONTAINERS FOR VENOUS BLOOD SPECIMEN COLLECTION' AND CLSI GP39-A6 REGULATIONS FOR 'EVACUATED TUBES AND ADDITIVES FOR BLOOD SPECIMEN COLLECTION'. TUBES WERE VERIFIED TO BE CORRECTLY ASSEMBLED AND HAD THE CORRECT FILL GUIDELINE POSITION. GREINER FILL MARK PROVIDES A VISUAL CONTROL OPPORTUNITY FOR THE PHLEBOTOMIST AND FOR THE LAB PERSONNEL TO CHECK FOR PROPER VOLUME COLLECTION OF SPECIMEN. BOTH STANDARDS SPECIFY THE DRAW VOLUME TO BE WITHIN +/- 10% RANGE OF THE NOMINAL FILL VOLUME. NO VISUAL DEVIATION IN FILL VOLUME, SPORADIC LOW OR HIGH FILL, COULD BE OBSERVED IN THE TESTED SAMPLES. ALL TUBES TESTED FILLED WITHIN THE +/-10% TOLERANCE RANGE. THE COMPLAINT COULD NOT BE DUPLICATED.
CUSTOMER STATES TUBES OVERFILLING. THEY STARTED KEEPING A LOG OF THE OVERFILLED TUBES AND NOTICED THEY WERE ALL THE SAME LOT NUMBER AND ALL WERE COMING FROM THE EMERGENCY ROOM. THEY RECEIVED ABOUT 20 SAMPLES OVER THE COURSE OF A WEEK FROM THE EMERGENCY AND ALL WERE OVERFILLED. COLLECTION BY SYRINGE WOULD BE UNLIKELY, AND MOST WERE PROBABLY STRAIGHT NEEDLE. THE LAB DID AN EXPERIMENT USING A URINE STRAW (URINE TRANSFER DEVICE) IN A CUP OF WATER TO ALLOW THE TUBES TO FILL NATURALLY AND THE TUBES ARE CLEARLY OVERFILLING. CUSTOMER PROVIDED PHOTOS FROM THEIR EXPERIMENT. CUSTOMER STATES THAT THE TYPE AND SCREENS ARE RUN ON THE IMMUCOR ECHO V2.0. THE INSTRUMENT GIVES ERROR PICK_FLUID (STATUS 5C00)-FLUID LEVEL TO HIGH ERROR. THE INSTRUMENT DOES NOT PRODUCE RESULTS AND WASTES REAGENTS. THE OVERFILLED TUBES WERE VISUALLY GREATER THAN 10% OVER THE FILL MARK AND CLEARLY VISIBLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 589151 | VACUETTE TUBE 6 ML K2E K2EDTA 13X100 PINK CAP-PINK RING, NON-RIDGED | EVACUATED BLOOD COLLECTION TUBE | JKA | GREINER BIO-ONE NA INC. | 456279 | B210233F |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |