FDA Adverse Event Injury Summary report: N

MID-C 105L

MDR report key: 11683216 · Received April 18, 2021

Report

Report Number
3013461531-2021-00018
Event Type
Injury
Date Received
April 18, 2021
Date of Event
March 9, 2021
Report Date
April 12, 2021
Manufacturer
APIFIX LTD
Product Code
QGP
PMA / PMN Number
H170001
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PO
Reporter Occupation
501

Narratives

Additional Manufacturer Narrative · 1

FIRST EVALUATION: PRODUCTION PROCESS: A REVIEW OF THE DHR DEMONSTRATED THAT THE MID-C SYSTEM WAS MANUFACTURED, TESTED, AND RELEASED ACCORDING TO APIFIX SPECIFICATIONS. PATIENT INFORMATION : BASED ON THE DISTRIBUTER INFORMATION THE REMOVAL SURGERY WAS PERFORMED DUE TO AN INFECTION. CORRECTIVE ACTION: FOLLOWING THE 2019 POST MARKET SURVELIANCE REPORT (DMS- (B)(4)), THE COMPANY DECIDED TO OPEN CAPA (B)(4) TO ADDRESS AN INCREASE IN LATE INFECTION COMPLAINTS. AS PART OF THE CAPA THE COMPANY: PERFORMED A VAST LITERATURE SEARCH TO IDENTIFY THE LATE INFECTION RATE REPORTED IN THE LITERATURE (DMS-(B)(4)). INVESTIGATED PRODUCTION LOTS ASSOCIATED WITH LATE INFECTION. THE COMPARED ANTIBIOTIC PROTOCOL USED BETWEEN CENTERS. RECEIVED CONSULTATION BY AN INFECTION SPECIALIST. THE CAPA INVESTIGATION DID NOT IDENTIFY A MANUFACTURING PROBLEM WITH IMPLANTS USED IN PATIENTS THAT DEVELOPED A LATE INFECTION AND THE LATE INFECTION RATE NOTED WAS WITHIN THE RATE REPORTED IN THE LITERATURE. THE COMPANY DID HOWEVER IDENTIFY THAT SITES APPLYING VANCOMYCIN POWDER BEFORE WOUND CLOSURE HAVE A SIGNIFICANTLY LOWER RATE OF LATE INFECTION COMPLAINTS. THE USE OF VANCOMYCIN POWDER WAS ALSO SUPPORTED BY THE INFECTION SPECIALIST CONSULTATION (DMS-(B)(4)) THE COMPANY COMMUNICATED THESE FINDINGS OF APPLYING VANCOMYCIN POWDER BEFORE WOUND CLOSURE TO THE SURGEONS (DMS-(B)(4)) AND IS CONTINUING TO FOLLOW THE RATE OF LATE INFECTION COMPLAINTS. RISK ASSESSMENT: LATE-ONSET INFECTION IS A KNOWN RISK THAT WAS ASSESSED AND RECORDED BY THE PRODUCT RISK ASSESSMENT DMS-(B)(4) REV Q HAZARD ID 21.5 THIS COMPLAINT DOES NOT CHANGE THE OCCURRENCES RATE. WE ASSUME THAT IN THE PRESENT CASE VANCOMYCIN WAS NOT APPLIED. THE TOTAL RATE OF LATE INFECTION IS 3.96% (N=20). OF THEM, ONLY 1 CASES WAS WHEN VANCOMYCIN WAS APPLIED WHICH IS 0.2% (1/505). THE RATE IS WELL WITHIN THE LITERATURE (0.2%-12.5% ,CER 727 REV T). THE COMPANY IS CONTINUING TO MONITOR LATE-INFECTION COMPLAINTS.

Description of Event or Problem · 1

THE PATIENT HAD A REMOVAL SURGERY ON (B)(6) 2021 DUE TO AN INFECTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
580958 MID-C 105L POSTERIOR RATCHETING ROD SYSTEM,,, QGP APIFIX LTD MID-C 105 AF 11-04-17

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention