ID NOW COVID-19 ASSAY
Report
- Report Number
- 1221359-2021-01006
- Event Type
- Malfunction
- Date Received
- April 16, 2021
- Date of Event
- March 12, 2021
- Report Date
- May 27, 2021
- Manufacturer
- ABBOTT DAIGNOSTICS SCARBOROUGH, INC
- Product Code
- QJR
- UDI-DI
- 10811877011269
- PMA / PMN Number
- EUA2000074
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HI, US
- Reporter Occupation
- 501
Narratives
THIS INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE PROVIDED. REFERENCE MFR REPORTS: 1221359-2021-00872, 1221359-2021-01007, 1221359-2021-01008.
ADDITIONAL INFORMATION: TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS (B)(4), INC. ON RETAINED KIT LOT 1023084 WITH INTERNAL POSITIVE QUALITY CONTROL. SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-000 / LOT: 1023084, TEST BASE PART NUMBER 190-430 / LOT: 1023084. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 1023084 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS (B)(4) WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER IT COULD POSSIBLY BE RELATED TO THE SPECIFIC PATIENT SAMPLE.
THE CUSTOMER REPORTED FALSE POSITIVES RESULTS WITH THE ID NOW COVID-19 ASSAY FOR MULTIPLE PATIENTS ON MULTIPLE DATES. THIS MFR REPORT ADDRESSES PATIENT 2 OF 4. FROM (B)(6) 2021 TO (B)(6) 2021 WE HAD ANOTHER 4 FALSE POSITIVE ID BASED ON TESTING 2 OR 3 OTHER SYSTEMS FROM PATIENTS BEING ADMITTED FOR CONDITIONS UNRELATED TO COVID-19. THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH THE ID NOW COVID-19 ASSAY PERFORMED ON (B)(6) 2021 ON A DIRECT NASOPHARYNGEAL KITTED SWAB. CONFIRMATION TESTING WAS PERFORMED ON THE SAME DAY USING LIAT, BD MAX. THIS GENERATED NEGATIVE RESULTS. THE CUSTOMER STATED THIS IS PRE-ADMISSION TESTING ON PATIENTS WHO WERE SUSPECTED TO HAVE COVID-19. PER THE CUSTOMER THE PATIENT WAS ASYMPTOMATIC. THE CUSTOMER CONFIRMED THERE WAS NO PATIENT HARM DUE TO TEST RESULTS. ADDITIONALLY, THE CUSTOMER CONFIRMED THERE WAS A DELAY AND IMPACT OF TREATMENT DUE TO TEST RESULTS. HOWEVER, NO ADDITIONAL INFORMATION WAS PROVIDED OTHER THAN "INACCURATELY CHARACTERIZED AS COVID POSITIVE." PER THE CUSTOMERS, THESE CASES WERE REPORTED TO US BY INFECTION CONTROL BECAUSE SOME OF THE POTENTIAL ISSUES WITH FALSE POSITIVES INCLUDING COHORTING PATIENTS WITH TRUE POSITIVE PATIENTS, HOSPITAL STAFF CONTACT INVESTIGATIONS, UNNECESSARY STAFF TESTING, DELAY/DISRUPTION IN TREATMENT FOR THE NON-COVID CONDITIONS ETC.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 574809 | ID NOW COVID-19 ASSAY | MOLECULAR IVD FOR ID NOW COVID-19 | QJR | ABBOTT DAIGNOSTICS SCARBOROUGH, INC | 1023084 | 10811877011269 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |