FDA Adverse Event Death Summary report: N

HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP

MDR report key: 11677948 · Received April 16, 2021

Report

Report Number
3007042319-2021-03219
Event Type
Death
Date Received
April 16, 2021
Date of Event
October 15, 2020
Report Date
June 8, 2021
Manufacturer
HEARTWARE, INC.
Product Code
DSQ
PMA / PMN Number
P100047
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

CORRECTIONS: D4, H4 A SUPPLEMENTAL REPORT IS BEING SUBMITTED FOR CORRECTIONS AND INVESTIGATION COMPLETION. D4 EXPIRATION DATE CORRECTED TO 28-FEB-2018 A ND UNIQUE IDENTIFIER (UDI) CORRECTED TO (B)(4). H4 DEVICE MANUFACTURE DATE CORRECTED TO 29-FEB-2016. PRODUCT EVENT SUMMARY: THE VENTRICULAR ASSIST DEVICE (VAD) (B)(6) WAS NOT RETURNED FOR EVALUATION. THIS COMPLAINT IS ASSOCIATED WITH A CLINICAL ADVERSE EVENT. INFORMATION RECEIVED FROM THE SITE INDICATED THAT THE PATIENT PRESENTED WITH CONFUSION AND DECREASED LOSS OF CONSCIOUSNESS, AND A BRAIN/HEAD COMPUTERIZED TOMOGRAPHY (CT) SCAN REVEALED AN EMBOLIC STROKE WHICH CONVERTED TO A HEMORRHAGIC STROKE. CARE WAS WITHDRAWN AND THE PATIENT SUBSEQUENTLY EXPIRED. BASED ON THE LIMITED INFORMATION AVAILABLE, THERE IS NO EVIDENCE TO INDICATE THAT A DEVICE MALFUNCTION OR PERFORMANCE ISSUE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. PER THE VAD INSTRUCTIONS FOR USE, NEUROLOGICAL DYSFUNCTION IS A KNOWN POTENTIAL COMPLICATION ASSOCIATED WITH THE IMPLANTATION OF A VAD. THERE WAS NO EVIDENCE THAT THE PATIENT HAD A HISTORY OF NEUROLOGICAL DYSFUNCTION EVENTS. POSSIBLE CLINICAL FACTORS THAT MAY HAVE CONTRIBUTED TO THIS EVENT INCLUDE THE PATIENT¿S PRE-EXISTING HISTORY AND RELATED COMORBIDITIES, THE PROGRESSION OF THEIR UNDERLYING DISEASE, ISSUES RELATED TO THE THERAPEUTIC USE OF ANTICOAGULANT AND ANTIPLATELET MEDICATIONS AND THE PATIENT'S COMPLEX POST-OPERATIVE COURSE. THERE ARE POSSIBLE PATIENT, PHARMACOLOGICAL AND PROCEDURAL FACTORS THAT MAY HAVE CONTRIBUTED TO THIS EVENT. INVESTIGATION OF THIS EVENT IS COMPLETED AND THE FILE WILL BE CLOSED. IF NEW INFORMATION IS RECEIVED, THE FILE WILL BE RE-OPENED AND A SUPPLEMENTAL WILL BE SUBMITTED. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

INVESTIGATION OF THIS EVENT IS PENDING AND A SUPPLEMENTAL REPORT WILL BE SENT UPON ITS COMPLETION. CONCOMITANT PRODUCTS: 5076-52 LEAD, IMPLANTED: (B)(6) 2012. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT PRESENTED WITH CONFUSION AND DECREASED LOSS OF CONSCIOUSNESS. THE PATIENT WAS ADMITTED, AND BRAIN/HEAD COMPUTERIZED TOMOGRAPHY (CT) SHOWED EMBOLIC STROKE WHICH CONVERTED TO A HEMORRHAGIC STROKE. IT WAS NOTED THAT THE PATIENT¿S INTERNATIONAL NORMALIZED RATIO (INR) WAS THERAPEUTIC AT THE TIME. THE STROKE WAS DEEMED LARGE, AND THE PATIENT¿S FAMILY DECIDE TO WITHDRAW CARE. IT WAS FURTHER REPORTED THAT THE VENTRICULAR ASSIST DEVICE (VAD) PATIENT SUBSEQUENTLY EXPIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
575293 HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP VENTRICULAR (ASSISST) BYPASS DSQ HEARTWARE, INC. 1103

Patients

Seq Age Sex Outcome Treatment
1 66 YR Death DDMB1D4 ICD, 6935M62 LEAD| DDMB1D4 ICD, 6935M62 LEAD