FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19 ASSAY

MDR report key: 11665026 · Received April 14, 2021

Report

Report Number
1221359-2021-00374
Event Type
Malfunction
Date Received
April 14, 2021
Date of Event
September 27, 2020
Report Date
April 14, 2021
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH,INC.
Product Code
QJR
UDI-DI
10811877011269
PMA / PMN Number
EUA 2000074
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS (B)(4) , INC. ON RETAINED KIT LOT 1003432 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST BASE PART NUMBER 190-430 / LOT: 1003432. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 1003432 SHOWED THAT THE COMPLAINT RATE IS (B)(4) . ABBOTT DIAGNOSTICS (B)(4) WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER IT COULD POSSIBLY BE RELATED TO THE SPECIFIC PATIENT SAMPLE. THE AVAILABLE EVIDENCE SUGGESTS THAT THIS DEVICE LOT IS PERFORMING WITHIN THE STATEMENTS MADE WITHIN PACKAGE INSERT RELATED TO % TEST AGREEMENT WITH THE EXPECTED RESULTS BY SAMPLE CONCENTRATION.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A FALSE POSITIVE TEST RESULT WITH TESTING OF IDNOW COVID-19 TEST ASSAY PERFORMED ON (B)(6) 2020. REPEAT TESTING WAS NOT PERFORMED. CONFIRMATION TESTING WITH PCR BIOFIRE GENERATED NEGATIVE RESULTS (CT RESULTS NOT PROVIDED). THE CUSTOMER CONFIRMED THERE WAS NO PATIENT HARM DUE TO THE TEST RESULTS. ADDITIONALLY, CUSTOMER REPORTED THERE WAS NO IMPACT OR DELAY IN PATIENT TREATMENT DUE TO THE TEST RESULTS. CUSTOMER REPORTED THAT PATIENT WAS SYMPTOMATIC, BUT NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED. POSITIVE RESULTS ARE INDICATIVE OF THE PRESENCE OF SARS-COV-2 RNA; CLINICAL CORRELATION WITH PATIENT HISTORY AND OTHER DIAGNOSTIC INFORMATION IS NECESSARY TO DETERMINE PATIENT INFECTION STATUS. POSITIVE RESULTS DO NOT RULE OUT BACTERIAL INFECTION OR CO-INFECTION WITH OTHER VIRUSES. DUE TO THE HIGH SENSITIVITY OF THE ASSAYS RUN ON THE INSTRUMENT, CONTAMINATION OF THE WORK AREA WITH PREVIOUS POSITIVE SAMPLES MAY CAUSE FALSE POSITIVE RESULTS. HANDLE SAMPLES ACCORDING TO STANDARD LABORATORY PRACTICES. CLEAN INSTRUMENTS AND SURROUNDING SURFACES ACCORDING TO INSTRUCTIONS PROVIDED IN THE CLEANING SECTION OF THE INSTRUMENT USER MANUAL. PER (B)(4) SCR REVE.0 POINT 1, DUE TO THE RISK OF A FALSE POSITIVE RESULT LEADING TO POSSIBLE EXPOSURE TO COVID-19 THROUGH THE QUARANTINE OF THE FALSE POSITIVE PATIENT WITH TRUE POSITIVE PATIENT(S), DELAYED EMERGENCY TREATMENT AND/OR INAPPROPRIATE TREATMENT WITH ANTIVIRAL THERAPY, THE INCIDENT SHALL BE CONSIDERED REPORTABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
563399 ID NOW COVID-19 ASSAY MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTICS SCARBOROUGH,INC. 1003432 10811877011269

Patients

Seq Age Sex Outcome Treatment
1