SPECTRUM INFUSION PUMP
Report
- Report Number
- 1314492-2021-01250
- Event Type
- Malfunction
- Date Received
- April 14, 2021
- Report Date
- April 14, 2021
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FRN
- PMA / PMN Number
- K133801
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THE DEVICE WAS RECEIVED FOR EVALUATION. VISUAL INSPECTION DID NOT IDENTIFY ANY ABNORMALITIES THAT COULD HAVE CONTRIBUTED TO THE REPORTED CONDITION. A SERVICE HISTORY REVIEW REVEALED NO INDICATION THAT THE PARTS REPLACED DURING SERVICING CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. EVALUATION INSPECTED THE DEVICE AND PERFORMED FLOW ACCURACY TESTING, WHICH FOUND THE DEVICE TO DELIVER WITHIN SPECIFICATION. THE REPORTED CONDITION WAS NOT VERIFIED. NO CORRECTION IS REQUIRED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A SPECTRUM PUMP RATE WOULD NOT GO ABOVE 252-254, VERY LOW. THE PROCESS STEP WAS NOT PROVIDED BY THE REPORTER. THERE WAS NO KNOWN PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.