TRUE METRIX AIR
Report
- Report Number
- 1000113657-2021-00253
- Event Type
- Malfunction
- Date Received
- April 13, 2021
- Date of Event
- March 17, 2021
- Report Date
- May 17, 2021
- Manufacturer
- TRIVIDIA HEALTH INC
- Product Code
- NBW
- UDI-DI
- 00021292007836
- PMA / PMN Number
- K150052
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NH, US
- Reporter Occupation
- OTHER
Narratives
SECTIONS WITH ADDITIONAL INFORMATION AS OF (B)(6)2021: D8: WAS THIS DEVICE SERVICED BY A THIRD PARTY. D9: DEVICE AVAILABLE FOR EVALUATION. H2 DEVICE EVALUATED BY MANUFACTURER. H6: UPDATED FDA'S TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, AND INVESTIGATION CONCLUSIONS H10: METER AND STRIPS WERE RETURNED FOR EVALUATION. REPORTED DEFECT NOT REPRODUCED.- NO DEFECT FOUND MOST LIKELY UNDERLYING ROOT MLC-058 CAUSE USER HAD AN INACCURATE REFERENCE: SELF: THE PERSON IS USING THEMSELVES AS THE REFERENCE OR HOW THEY FEEL AT THE TIME THEY RUN THE BLOOD TEST.
INTERNAL REPORT REFERENCE NUMBER: (B)(4). METER AND TEST STRIPS WERE NOT RETURNED FOR EVALUATION. NOTE: MANUFACTURER CONTACTED CUSTOMER IN A FOLLOW-UP CALL ON 31-MAR-2021 TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - ABLE TO ESTABLISH CONTACT WITH CUSTOMER WHO STATED REPLACEMENT PRODUCTS RESOLVED INITIAL CONCERN.
CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 243, 315, 250, 335 AND 326MG/DL. THE CUSTOMER¿S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 80-120MG/DL AND EXPECTED RESULT TWO HOURS POST MEAL IS 150MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BLOOD TEST WAS PERFORMED BY THE CUSTOMER FASTING AND PRODUCED TEST RESULT OF 230MG/DL USING TRUE METRIX AIR METER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 04/28/2022 AND TEST STRIPS WERE OPENED A FEW DAYS PRIOR TO CALL. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (B)(6).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 557998 | TRUE METRIX AIR | SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER | NBW | TRIVIDIA HEALTH INC | STRIP, TRUE METRIX NFRS24/CS 50CT MG/DL | MX4418S | 00021292007836 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |