SYRINGE 30ML LL S/C 56
Report
- Report Number
- 1911916-2021-00309
- Event Type
- Malfunction
- Date Received
- April 13, 2021
- Date of Event
- March 15, 2021
- Report Date
- April 2, 2021
- Manufacturer
- BD MEDICAL (BD WEST) MEDICAL SURGICAL
- Product Code
- FMF
- PMA / PMN Number
- K980987
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. MEDICAL DEVICE LOT #: AN INVALID LOT # OF 0163209 WAS PROVIDED BY THE INITIAL REPORTER: DEVICE EXPIRATION DATE: UNKNOWN. DEVICE MANUFACTURE DATE: UNKNOWN. INVESTIGATION SUMMARY: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 302832, LOT NUMBER 0163029. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION OF THIS LOT THAT COULD HAVE CONTRIBUTED TO THIS REPORTED DEFECT. TO AID IN THE INVESTIGATION, ONE PHYSICAL SAMPLE AND ONE PHOTO WERE PROVIDED FOR EVALUATION BY OUR QUALITY TEAM. THE SAMPLE CAME IN A SEALED PACKAGING BLISTER AND A VISUAL INSPECTION WAS PERFORMED. THE SAMPLE HAS A SCALE PRINTING ISSUE. THE STOPPER HAS NO DAMAGES OR IMPERFECTIONS, BUT THE BARREL DOES HAVE A SCRATCH AT THE BOTTOM. THE PHOTO PROVIDED SHOWS THE SAMPLE RECEIVED. THE SCALE PRINTING DEFECT CAN OCCUR IF THERE IS A JAM DURING PRINTING INDUCING THE SYMPTOM REPORTED. IT IS ALSO POSSIBLE FOR A JAM DURING MANUFACTURING TO INDUCE THE SCRATCH AT THE BOTTOM PART OF THE SYRINGE BARREL. BASED ON THE INVESTIGATION WITH THE RETURNED SAMPLE AND PHOTO SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER IS CONFIRMED. FURTHER ACTION HAS NOT BEEN DETERMINED NECESSARY AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR QUALITY TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. INVESTIGATION CONCLUSION: BASED ON THE INVESTIGATION AND WITH THE SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER IS CONFIRMED. WE WILL CONTINUE MONITORING THE COMPLAINT TREND FOR THE PRODUCT AND SYMPTOM. PROBABLE ROOT CAUSE. IT COULD BE POSSIBLE THAT A JAM OCCURRED AT THE SCALE PRINTING INDUCING THE SYMPTOM REPORTED AND THE SCRATCH AT THE BOTTOM OF THE SYRINGE BARREL.
IT WAS REPORTED THAT THE SYRINGE 30ML LL S/C 56 EXPERIENCED SCALE MARKING ISSUES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: POOR SCALE MARKER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 553009 | SYRINGE 30ML LL S/C 56 | PISTON SYRINGE | FMF | BD MEDICAL (BD WEST) MEDICAL SURGICAL | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |