FDA Adverse Event Malfunction Summary report: N

SYRINGE 30ML LL S/C 56

MDR report key: 11657204 · Received April 13, 2021

Report

Report Number
1911916-2021-00309
Event Type
Malfunction
Date Received
April 13, 2021
Date of Event
March 15, 2021
Report Date
April 2, 2021
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
FMF
PMA / PMN Number
K980987
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. MEDICAL DEVICE LOT #: AN INVALID LOT # OF 0163209 WAS PROVIDED BY THE INITIAL REPORTER: DEVICE EXPIRATION DATE: UNKNOWN. DEVICE MANUFACTURE DATE: UNKNOWN. INVESTIGATION SUMMARY: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER 302832, LOT NUMBER 0163029. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION OF THIS LOT THAT COULD HAVE CONTRIBUTED TO THIS REPORTED DEFECT. TO AID IN THE INVESTIGATION, ONE PHYSICAL SAMPLE AND ONE PHOTO WERE PROVIDED FOR EVALUATION BY OUR QUALITY TEAM. THE SAMPLE CAME IN A SEALED PACKAGING BLISTER AND A VISUAL INSPECTION WAS PERFORMED. THE SAMPLE HAS A SCALE PRINTING ISSUE. THE STOPPER HAS NO DAMAGES OR IMPERFECTIONS, BUT THE BARREL DOES HAVE A SCRATCH AT THE BOTTOM. THE PHOTO PROVIDED SHOWS THE SAMPLE RECEIVED. THE SCALE PRINTING DEFECT CAN OCCUR IF THERE IS A JAM DURING PRINTING INDUCING THE SYMPTOM REPORTED. IT IS ALSO POSSIBLE FOR A JAM DURING MANUFACTURING TO INDUCE THE SCRATCH AT THE BOTTOM PART OF THE SYRINGE BARREL. BASED ON THE INVESTIGATION WITH THE RETURNED SAMPLE AND PHOTO SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER IS CONFIRMED. FURTHER ACTION HAS NOT BEEN DETERMINED NECESSARY AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR QUALITY TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. INVESTIGATION CONCLUSION: BASED ON THE INVESTIGATION AND WITH THE SAMPLE ANALYSIS THE SYMPTOM REPORTED BY THE CUSTOMER IS CONFIRMED. WE WILL CONTINUE MONITORING THE COMPLAINT TREND FOR THE PRODUCT AND SYMPTOM. PROBABLE ROOT CAUSE. IT COULD BE POSSIBLE THAT A JAM OCCURRED AT THE SCALE PRINTING INDUCING THE SYMPTOM REPORTED AND THE SCRATCH AT THE BOTTOM OF THE SYRINGE BARREL.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SYRINGE 30ML LL S/C 56 EXPERIENCED SCALE MARKING ISSUES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: POOR SCALE MARKER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
553009 SYRINGE 30ML LL S/C 56 PISTON SYRINGE FMF BD MEDICAL (BD WEST) MEDICAL SURGICAL UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1