ABBOTT PRISM HCV ASSAY KIT
Report
- Report Number
- 1415939-2021-00014
- Event Type
- Malfunction
- Date Received
- April 13, 2021
- Date of Event
- February 26, 2021
- Report Date
- June 9, 2021
- Manufacturer
- ABBOTT LABORATORIES
- Product Code
- QHM
- UDI-DI
- 00380740062996
- PMA / PMN Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). WAS THIS DEVICE SERVICED BY A THIRD PARTY? NO. (B)(6). ALL AVAILABLE PATIENT INFORMATION INCLUDED. NO ADDITIONAL INFORMATION WAS PROVIDED. AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE THIS MDR IS BEING SUBMITTED FOR THE PRISM (B)(6) REAGENT, LIST 06D18-68
A REVIEW OF COMPLAINTS ABBOTT PRISM CHAGAS, LIST NUMBER (LN) (B)(4), LOT 15094M700, ABBOTT PRISM HCV, (B)(4), LOT 20049M700 , AND ABBOTT PRISM HTLV-I/HTLV-II, (B)(4), LOT 22075M700 IDENTIFIED NORMAL COMPLAINT ACTIVITY FOR ALL LOTS INVOLVED. A REVIEW OF TICKET TRENDING WAS PERFORMED AN NO TREND WAS IDENTIFIED FOR THE CUSTOMER COMPLAINT FOR THE ABBOTT PRISM CHAGAS, LIST NUMBER (LN) (B)(4)), ABBOTT PRISM HCV, (B)(4), AND THE ABBOTT PRISM HTLV-I/HTLV-II, (B)(4) REAGENTS. A REVIEW OF CUSTOMER DATA WAS PERFORMED INCLUDING THE REPEAT REACTIVE RATES (RRR) FOR ABBOTT PRISM CHAGAS LOT 15094M700, ABBOTT PRISM HCV LOT 20049M700 AND ABBOTT PRISM HTLV-I/HTLV-II LOT 22075M700. IT WAS FOUND THAT THE CUSTOMER RRR WERE WITHIN THE PACKAGE INSERT RANGES FOR DATA FOR ABBOTT PRISM CHAGAS LOT 15094M700 AND ABBOTT PRISM HCV LOT 20049M700. FOR ABBOTT PRISM HTLV-I/HTLV-II LOT 22075M700, THE (B)(4) RRR EXCEEDED THE OVERALL TOTAL DONOR UPPER 95% CONFIDENCE LIMIT BUT WAS STILL WITHIN THE RANGE OF EXPECTED PERFORMANCE FOR ANY SINGLE ABBOTT PRISM HTLV-I/HTLV-II LOT ((B)(4)) AS SHOWN IN THE PACKAGE INSERT. MANUFACTURING DOCUMENTATION WAS REVIEWED AND NO ISSUES WERE IDENTIFIED. ADDITIONALLY, LABELING WAS REVIEWED AND SUFFICIENTLY ADDRESSES THE CUSTOMER ISSUE. BASED ON THE RESULTS OF THIS INVESTIGATION, THERE IS NO SYSTEMIC ISSUE OR DEFICIENCY WAS IDENTIFIED FOR THE PRISM CHAGAS LOT 15094M700, PRISM HCV LOT 20049M700 AND PRISM HTLV-I/HTLV-II LOT 22075M700. THIS FOLLOW UP IS SUBMITTED TO POPULATE FIELDS D8 AND/OR H6 WITH DATA THAT HAD PREVIOUSLY BEEN PROVIDED IN FIELD H10. THERE IS NO CHANGE TO THE CONTENT OF THE DATA.
THE CUSTOMER REPORTED (B)(6) PRISM CHAGAS, PRISM (B)(6), AND PRISM HTLV RESULTS WERE GENERATED. THE CUSTOMER STATED THAT THE SAMPLES WERE ARCHIVED THEN RUN ON THE ALINITY S FOR VALIDATION TESTING AS ¿KNOWN¿ (B)(6) BUT THEY GENERATED (B)(6) RESULTS. THE FOLLOWING RESULTS WERE PROVIDED: SID (B)(6)- (B)(6) 2021 PRISM CHAGAS REPEAT PRISM CHAGAS REPEAT REACTIVE (1.40, 1.47, 1.39 S/CO) (LOT 15094M700); (B)(6) 2021- ALINITY S NON-REACTIVE (0.14 S/CO); ESA INDETERMINATE SID (B)(6)- (B)(6) 2021 PRISM CHAGAS REPEAT REACTIVE (1.78, 2.61, 2.55 S/CO) (LOT 15094M700); (B)(6) 2021- ALINITY S NON-REACTIVE (0.02 S/CO); ESA NEGATIVE SID (B)(6) - (B)(6) 2021 PRISM HTLV REPEAT REACTIVE (1.94, 1.88, 1.76 S/CO);(B)(6) 2021- ALINITY S NON-REACTIVE (0.14 S/CO); MP HTLV BLOT NEGATIVE SID (B)(6)- (B)(6) 2021 PRISM HTLV REPEAT REACTIVE (1.93, 2.01, AND 1.99 S/CO); (B)(6) 2021- ALINITY S NON-REACTIVE (0.13 S/CO); MP HTLV BLOT INDETERMINATE SID (B)(6)- (B)(6) 2021 PRISM HTLV REPEAT REACTIVE (1.25, 1.47, 1.39 S/CO); (B)(6) 2021- ALINITY S NON-REACTIVE (0.16 S/CO); MP HTLV BLOT INVALID SID (B)(6)- (B)(6) 2021 PRISM (B)(6) REPEAT (B)(6); (B)(6) 2021- ALINITY S (B)(6); ORTHO (B)(6), ROCHE NAT (B)(6) THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
THE CUSTOMER REPORTED FALSE REPEAT REACTIVE PRISM CHAGAS, PRISM HCV, AND PRISM HTLV RESULTS WERE GENERATED. THE CUSTOMER STATED THAT THE SAMPLES WERE ARCHIVED THEN RUN ON THE ALINITY S FOR VALIDATION TESTING AS ¿KNOWN¿ POSITIVES BUT THEY GENERATED NON-REACTIVE RESULTS. THE FOLLOWING RESULTS WERE PROVIDED: SID (B)(4)- (B)(6) 2021 PRISM CHAGAS REPEAT REACTIVE (1.40, 1.47, 1.39 S/CO) (LOT 15094M700); (B)(6) 2021- ALINITY S NON-REACTIVE (0.54 S/CO); ESA INDETERMINATE. SID (B)(4)- (B)(6) 2021 PRISM CHAGAS REPEAT REACTIVE (1.78, 2.61, 2.55 S/CO) (LOT 15094M700); (B)(6) 2021- ALINITY S NON-REACTIVE (0.02 S/CO); ESA NEGATIVE. SID (B)(4)- (B)(6) 2021 PRISM HTLV REPEAT REACTIVE (1.94, 1.88, 1.76 S/CO); (B)(6) 2021- ALINITY S NON-REACTIVE (0.14 S/CO); MP HTLV BLOT NEGATIVE. SID (B)(4)- (B)(6) 2021 PRISM HTLV REPEAT REACTIVE (1.93, 2.01, AND 1.99 S/CO); (B)(6) 2021- ALINITY S NON-REACTIVE (0.13 S/CO); MP HTLV BLOT INDETERMINATE. SID (B)(4)- (B)(6) 2021 PRISM HTLV REPEAT REACTIVE (1.25, 1.47, 1.39 S/CO); (B)(6) 2021- ALINITY S NON-REACTIVE (0.16 S/CO); MP HTLV BLOT INVALID. SID (B)(4)- (B)(6) 2021 PRISM HCV REPEAT REACTIVE (1.26, 1.20, 1.24 S/CO); (B)(6) 2021- ALINITY S NON-REACTIVE (0.06 S/CO); ORTHO HCV NONREACTIVE, ROCHE NAT NEGATIVE THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 557189 | ABBOTT PRISM HCV ASSAY KIT | HEPATITIS C VIRUS ENCODED ANTIGENS (RECOMBINANT C100-3, HCR43, NS5) | QHM | ABBOTT LABORATORIES | 20049M700 | 00380740062996 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | PRISM 6 CHANNEL-CE, 06A36-04, 1174| PRISM 6 CHANNEL-CE, 06A36-04, 1174| PRISM CHAGAS US A, 07K35-68, 15094M700| PRISM CHAGAS US A, 07K35-68, 15094M700| PRSM HTLV RGT KIT, 06E50-68, 22075M700| PRSM HTLV RGT KIT, 06E50-68, 22075M700 |