MEERA EU WITH AUTO DRIVE
Report
- Report Number
- 3013876692-2021-00025
- Event Type
- Injury
- Date Received
- April 13, 2021
- Date of Event
- April 4, 2021
- Report Date
- October 12, 2021
- Manufacturer
- HOLGER ULLRICH
- Product Code
- FQO
- Adverse Event
- Yes
- Report Source
- Distributor report
- Reporter Location
- EI
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
A GETINGE-MAQUET FIELD SERVICE ENGINEER HAS INVESTIGATED THE AFFECTED TABLE AND HAS NOT FOUND ANY DAMAGE OR MALFUNCTION. FURTHER INFORMATION CONCERNING THIS INCIDENT WAS REQUESTED FROM THE CLINIC, BUT NOT PROVIDED. WE ASSUME THAT THE FOLLOWING HAS HAPPENED. THE TABLES' DOWNWARD MOVEMENT WAS BLOCKED BY AN OBSTACLE, E.G. AN INSTRUMENT TABLE. WHEN THE TABLE OR A PART OF IT WAS MOVED DOWNWARDS, THE TABLES BASE WAS LIFTED OF THE FLOOR. ONCE THE OBSTACLE BROKE OR SLIPPED OUT, THE TABLES BASE WAS LOWERED BACK ON THE FLOOR AND CRUSHED THE DOCTORS FOOT. IN THE INSTRUCTIONS FOR USE (IFU) IT IS STATED THE USER HAS TO PAY ATTENTION AND AVOID COLLISIONS WHEN OPERATING THE TABLE. IN CHAPTER 2 OF THE IFU POSSIBLE HAZARDS ARE STATED: "WARNING! RISK OF INJURY! WHEN ADJUSTING AND MOVING THE OR TABLE, THE TABLE TOP OR THE ACCESSORIES, COLLISIONS MAY OCCUR WITH THE PATIENT, BETWEEN THE INDIVIDUAL PRODUCTS OR PARTS POINTING DOWNWARDS. DURING THE ADJUSTMENT PROCEDURES, ALWAYS PAY ATTENTION TO THE OR TABLE AND ACCESSORIES AND AVOID COLLISIONS. ENSURE THAT TUBES, CABLES AND DRAPES ARE NOT TRAPPED." GETINGE-MAQUET GMBH PROVIDES PRODUCT FAILURE INVESTIGATION, ANALYSIS AND RESOLUTION FOR THE DEVICE DESCRIBED IN THIS REPORT.
THE CLINIC STATED THAT THE TABLE WAS DRIVEN INTO THE TRENDELENBURG POSITION WHEN THE ISSUE OCCURRED. AN X-RAY WAS TAKEN OF THE INJURED FOOT. NO BROKEN BONES WERE FOUND DUE TO THIS ACCIDENT. THE SURGEON GOT SOME PAINKILLER. NO MAYOR DELAY OF THE SURGERY HAS OCCURRED. MANUFACTURER REFERENCE# (B)(4).
A GETINGE-MAQUET FIELD SERVICE ENGINEER HAS INVESTIGATED THE AFFECTED TABLE AND HAS NOT FOUND ANY DAMAGE OR MALFUNCTION. AT THE TIME OF THIS REPORT THE INVESTIGATION IS STILL ONGOING. WHEN THE INVESTIGATION IS COMPLETE THE REPORT WILL BE UPDATED AND A FOLLOW UP MEDWATCH WILL BE SUBMITTED.
THE FOLLOWING WAS REPORTED. THE PATIENT WAS ON THE TABLE WHEN THE TABLE WAS LOWERED AS REQUESTED. THE CUSTOMER DESCRIBED THAT THE TABLE "JOLTED" AND A PART OF THE TABLES BASE FELL ON THE LEFT FOOT OF A SURGEON. THIS CAUSED AN INJURY. WE HAVE REQUESTED FURTHER INFORMATION CONCERNING THE INCIDENT AND THE EXTENT OF THE INJURY, BUT NOT RECEIVED IT YET. MANUFACTURER REFERENCE# (B)(4).