COLLEAGUE VOLUMETRIC INFUSION PUMP
MDR report key: 1165564
·
Received September 22, 2008
Report
- Report Number
- 6000001-2007-02561
- Event Type
- Malfunction
- Date Received
- September 22, 2008
- Date of Event
- January 2, 2007
- Report Date
- January 2, 2007
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Removal / Correction Number
- CONTINUED ON H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
EVALUATION SUMMARY: THE DEVICE EVALUATION WAS COMPLETED AND FAILURE CODE 570:320:844:000, FOUND TO BE DUE TO THE DEPLETED (DAMAGED) BATTERY CONDITION, WAS CONFIRMED. SERVICE INSTALLED NEW BATTERIES AND TESTED THE DEVICE. A REVIEW OF THE COMPLAINT HISTORY REVEALED SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT AND THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED UNDER CAPA.
Description of Event or Problem · 1
THE DEVICE WAS RECEIVED FOR SERVICE. DURING SERVICE TESTING, FAILURE CODE 570:320:844:0000 WAS FOUND. ACCORDING TO THE HOSPITAL REPRESENTATIVE THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.