B.BRAUN
Report
- Report Number
- 9610825-2021-00117
- Event Type
- Malfunction
- Date Received
- April 12, 2021
- Date of Event
- February 26, 2021
- Report Date
- April 12, 2021
- Manufacturer
- B.BRAUN MELSUNGEN AG
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN (B)(6) AG INTERNAL REPORT (B)(4). REFERENCE SAMPLE: MODEL: SPACE COVER COMFORT. ARTICLE NO.: 8713145. SERIAL NO./LOT: 58603. PRODUCTION DATE: 2016-06-13. WARRANTY CLAIM: NO. RESULTS: A VISUAL TEST WAS PERFORMED ON THE SPACE COVER COMFORT. ALL COVER CAPS WERE PRESENT. THE HOUSING OF THE SPACE STATION WAS SOILED BY LIQUIDS. FOR A FUNCTION TEST THE SPACE COVER COMFORT WAS CONNECTED TO A SPACE STATION WITH INSERTED INFUSION PUMPS. THE SPACE COVER COMFORT FUNCTIONED ACCORDING TO SPECIFICATIONS. AFTER DISASSEMBLING THE SPACE COVER COMFORT, NO DAMAGES WERE FOUND INSIDE THE DEVICE. ASSESSMENT: THE COMPLAINT COULD NOT BE CONFIRMED. THE FIRE AT THE SPACE STATION DID NOT AFFECT OR DAMAGE THE SPACE COVER COMFORT. NOTE: THIS REPORT IS BEING FILED FOR AN ITEM NUMBER THAT IS NOT SOLD IN THE UNITED STATES, HOWEVER SIMILAR ITEMS ARE SOLD IN THE UNITED STATES BY B. BRAUN MEDICAL, INC.
AS REPORTED BY THE USER FACILITY INFORMATION BY BBM SALES ORGANIZATION IN (B)(6): FIRE AT SPACE STATION. SUDDENLY THERE WAS A FIRE AT THE DOCKING STATION OF THE SPACE INFUSION SYSTEM WITH CLEAR FLYING SPARKS. NOTE: EIGHT REPORTS ARE BEING FILED FOR ONE EVENT, BECAUSE ALL EIGHT DEVICES WERE INVOLVED IN THE ONE THERMAL EVENT AND THE CAUSE OF THE THERMAL EVENT COULD NOT BE TRACED TO ONE OF THE DEVICES. THIS REPORT IS BEING FILED FOR THE SPACESTATION COVER IN USE AT THE TIME OF THE EVENT. REPORT NUMBERS 9610825-2021-00110, 9610825-2021-00111, 9610825-2021-00112, 9610825-2021-00113, 9610825-2021-00114, 9610825-2021-00115, AND 9610825-2021-00116 ARE BEING FILED FOR THE OTHER DEVICES INVOLVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 548738 | B.BRAUN | PUMP, INFUSION | FRN | B.BRAUN MELSUNGEN AG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |