BD BBL CHOCOLATE II AGAR (GC II AGAR WITH HEMOGLOBIN AND ISOVITALEX)
Report
- Report Number
- 1119779-2021-00647
- Event Type
- Malfunction
- Date Received
- April 12, 2021
- Date of Event
- March 22, 2021
- Report Date
- September 23, 2021
- Manufacturer
- BECTON, DICKINSON & CO. (SPARKS)
- Product Code
- JTY
- UDI-DI
- 10382902212673
- PMA / PMN Number
- PREAMENDMENT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H.6. INVESTIGATION: DURING MANUFACTURING OF MATERIAL 221267, MEDIA IS FORMULATED AND SENT THROUGH A HIGH TEMPERATURE SHORT TIME STERILIZER TO REMOVE BIOBURDEN. THE PETRI DISHES ARE SUBJECTED TO UV RADIATION TO DECREASE BIOBURDEN. THE PETRI DISHES ARE FILLED IN A POSITIVE PRESSURE HEPA FILTERED ENVIRONMENT. THE FILLED PLATES ARE COOLED AND IMMEDIATELY WRAPPED INTO SLEEVES TO DECREASE THE INTRODUCTION OF MICROBES. SLEEVES ARE THEN PACKAGED INTO CARTONS AND THEN TRANSFERRED TO A REFRIGERATED TRUCK (2 TO 8 DEGREES C) FOR SHIPMENT TO THE DISTRIBUTOR. BD DISTRIBUTORS ARE PROVIDED WITH THE STORAGE GUIDELINES FOR THE SHIPPING AND HANDLING OF BD MEDIA OF 2 TO 8 DEGREES C IN A DARK PLACE. THE BATCH HISTORY RECORDS FOR BATCHES 1028365 AND 0329426 WERE SATISFACTORY AT TIME OF RELEASE AND NO QUALITY NOTIFICATIONS WERE GENERATED DURING MANUFACTURING AND INSPECTION. THE RELEASE TESTING THAT IS PERFORMED ON THIS PRODUCT DOES INCLUDE BIOBURDEN TESTING. A SAMPLE OF PLATES ARE INCUBATED AT 25 DEGREES C AND AT 35 DEGREES C FOR APPROXIMATELY 72 HOURS. ALL BIOBURDEN TESTING PERFORMED ON THIS BATCH WAS SATISFACTORY PER BD INTERNAL PROCEDURES. AFFECTED PRODUCT DOES NOT HAVE ANY STERILITY CLAIMS; THE PRODUCT IS TESTED FOR BIOBURDEN PRIOR TO RELEASE TO ENSURE THAT IT CONFORMS TO PRODUCT SPECIFICATIONS. HOWEVER, THIS DOES NOT ENSURE THAT THE END-USER WILL NOT RECEIVE A CONTAMINATED PLATE. THE COMPLAINT HISTORY WAS REVIEWED, AND NO OTHER COMPLAINTS HAVE BEEN TAKEN EITHER BATCH 1028365 OR BATCH 0329426. RETENTION SAMPLES FROM BATCHES 1028365 AND 0329426 WERE NOT AVAILABLE FOR INSPECTION. TWO PHOTOS WERE RECEIVED FOR INVESTIGATION. ONE PHOTO SHOWS THE AGAR SURFACE OF A PLATE WITH APPEARS TO BE A COLONY OF MICROBIAL GROWTH THAT HAS BEEN PICKED. THE OTHER PHOTO SHOWS THE BOTTOM OF A PLATE FROM BATCH 1028365 (TIME STAMP 2049) WITH THE PLATE PRINT FEATURED FOR BATCH VERIFICATION. NO RETURN SAMPLES WERE RECEIVED FOR INVESTIGATION. THE COMPLAINT HAS BEEN CONFIRMED FOR BATCH 1028365. THIS COMPLAINT CANNOT BE CONFIRMED FOR BATCH 0329426. BD WILL CONTINUE TO TREND COMPLAINTS FOR CONTAMINATION. DUE TO THE LOW DEFECT RATE FOR BATCH 1028365, NO ACTIONS ARE PLANNED AT THIS TIME. H3 OTHER TEXT : SEE H.10.
IT WAS REPORTED THAT PRIOR TO USE WITH BD BBL¿ CHOCOLATE II AGAR (GC II AGAR WITH HEMOGLOBIN AND ISOVITALEX¿) BACTERIAL CONTAMINATION WAS DISCOVERED. THIS OCCURRED WITH 10 PLATES OF LOT# 1028365 AND 30 WITH LOT# 0329426. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORTS CONTAMINATION IS BACTERIAL AND APPEARS TO BE A BACILLIUS BECAUSE OF A MUCOID APPEARANCE. LOT: 1028365 (B)(4). LOT: 0329426 (B)(4).
MULTIPLE LOT NUMBERS: THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR ADDITIONAL LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 1028365. MEDICAL DEVICE EXPIRATION DATE: 2021-05-25. DEVICE MANUFACTURE DATE: 2021-01-28. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT PRIOR TO USE WITH BD BBL¿ CHOCOLATE II AGAR (GC II AGAR WITH HEMOGLOBIN AND ISOVITALEX¿) BACTERIAL CONTAMINATION WAS DISCOVERED. THIS OCCURRED WITH 10 PLATES OF LOT# 1028365 AND 30 WITH LOT# 0329426. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER REPORTS CONTAMINATION IS BACTERIAL AND APPEARS TO BE A BACILLIUS BECAUSE OF A MUCOID APPEARANCE. LOT: 1028365, 10 AFFECTED OF 100 RECEIVED. LOT: 0329426, 30 AFFECTED OF 200 RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 545582 | BD BBL CHOCOLATE II AGAR (GC II AGAR WITH HEMOGLOBIN AND ISOVITALEX) | CULTURE MEDIA, FOR ISOLATION OF PATHOGENIC NEISSERIA | JTY | BECTON, DICKINSON & CO. (SPARKS) | 221267 | 0329426 | 10382902212673 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |