FDA Adverse Event Malfunction Summary report: N

RICHMAR

MDR report key: 11648064 · Received April 12, 2021

Report

Report Number
3012316249-2021-00016
Event Type
Malfunction
Date Received
April 12, 2021
Report Date
April 21, 2021
Manufacturer
COMPASS HEALTH BRANDS
Product Code
IMG
PMA / PMN Number
K032941
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 0

DEVICE RETURNED FOR INSPECTION: COMPLAINT CONFIRMED. THE FISCHER CONNECTOR WAS DAMAGED. THE FISCHER CONNECTOR WAS REPAIRED THEN THE APPLICATOR WAS TESTED AND PASSED. NO PARTS WERE NEEDED FOR REPAIRS. THE OUTER SHELL OF THE FISCHER CONNECTOR WAS AROUND THE WIRES WHEN WE GOT IT SO THEY WERE NOT VISIBLE AT FIRST BUT WHEN THE FISCHER CONNECTOR CASE WAS OPENED THE RED WIRE WAS PULLED OUT OF THE PINS SO THE END OF THE WIRE WAS VISIBLE.

Description of Event or Problem · 1

THE PLUG WAS PULLED OUT OF THE UNIT AND THE CONNECTOR BROKE OFF. THE CUSTOMER REPORTED THERE MAY BE ONE WIRE THAT IS SHOWING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
545759 RICHMAR APPLICATOR AUTOSOUND IMG COMPASS HEALTH BRANDS 410-200

Patients

Seq Age Sex Outcome Treatment
1