FDA Adverse Event
Malfunction
Summary report: N
RICHMAR
MDR report key: 11648064
·
Received April 12, 2021
Report
- Report Number
- 3012316249-2021-00016
- Event Type
- Malfunction
- Date Received
- April 12, 2021
- Report Date
- April 21, 2021
- Manufacturer
- COMPASS HEALTH BRANDS
- Product Code
- IMG
- PMA / PMN Number
- K032941
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 0
DEVICE RETURNED FOR INSPECTION: COMPLAINT CONFIRMED. THE FISCHER CONNECTOR WAS DAMAGED. THE FISCHER CONNECTOR WAS REPAIRED THEN THE APPLICATOR WAS TESTED AND PASSED. NO PARTS WERE NEEDED FOR REPAIRS. THE OUTER SHELL OF THE FISCHER CONNECTOR WAS AROUND THE WIRES WHEN WE GOT IT SO THEY WERE NOT VISIBLE AT FIRST BUT WHEN THE FISCHER CONNECTOR CASE WAS OPENED THE RED WIRE WAS PULLED OUT OF THE PINS SO THE END OF THE WIRE WAS VISIBLE.
Description of Event or Problem · 1
THE PLUG WAS PULLED OUT OF THE UNIT AND THE CONNECTOR BROKE OFF. THE CUSTOMER REPORTED THERE MAY BE ONE WIRE THAT IS SHOWING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 545759 | RICHMAR | APPLICATOR AUTOSOUND | IMG | COMPASS HEALTH BRANDS | 410-200 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |