EVIS EXERA II ULTRASONIC BRONCHOFIBERVIDEOSCOPE
Report
- Report Number
- 8010047-2021-04822
- Event Type
- Malfunction
- Date Received
- April 12, 2021
- Date of Event
- March 17, 2021
- Report Date
- November 16, 2021
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORP.
- Product Code
- PSV
- UDI-DI
- 04953170356360
- PMA / PMN Number
- K070983
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- 501
Narratives
THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO CORRECT THE " TYPE OF REPORTABLE EVENT." DURING THE REMEDIAL REVIEW OF THIS COMPLAINT IT WAS DISCOVERED THAT A DATA ENTRY ERROR IN SECTION H1 OCCURRED. AS THERE WAS NO PATIENT INVOLVEMENT OR DEATH REPORTED BY THE CUSTOMER, THE TYPE OF REPORTABLE EVENT HAS BEEN DETERMINED TO BE A MALFUNCTION.
AS PART OF OUR INVESTIGATION, WHILE ONSITE THE ESS COMPLETED A REPROCESSING IN-SERVICE WITH STAFF. THE ESS RECOMMENDED THAT THE CUSTOMER FOLLOW ALL OLYMPUS REPROCESSING PROCEDURES AS DOCUMENTED IN THE ONTRACK FORMS AND REPROCESSING MANUALS. THE SCOPE WILL NOT BE RETURNED FOR EVALUATION. IN ADDITION, THE LEGAL MANUFACTURER REVIEWED THE CONTENT OF THE COMPLAINT. THE LEGAL MANUFACTURER PERFORMED A DHR REVIEW AND CONFIRMED THE SUBJECT DEVICE WAS SHIPPED IN ACCORDANCE WITH SPECIFICATIONS. THE ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED. HOWEVER, THE LEGAL MANUFACTURER PRESUMED THAT HUMAN ERROR MAY HAVE RESULTED IN FAILURE OF FACILITY TO PERFORM ADEQUATE SCOPIC IRRIGATION (BALLOON CHANNEL BRUSHING). ULTRASONIC BRONCHOFIBERVIDEOSCOPE BF-UC180F CHAPTER 7 CLEANING, DISINFECTION, AND STERILIZATION PROCEDURES CHANNEL CLEANING BRUSH (BW-7B/REUSABLE); SINGLE USE SINGLE-ENDED CLEANING BRUSH (BW-400B); THE CHANNEL CLEANING BRUSH OR THE SINGLE USE SINGLE-ENDED CLEANING BRUSH IS USED TO CLEAN THE BALLOON CHANNEL CLEANING THE BALLOON CHANNEL IN THE INSERTION SECTION (LOCATION E). GRIP THE CHANNEL CLEANING BRUSH (BW-7B) OR THE SINGLE USE SINGLE-ENDED CLEANING BRUSH (BW-400B) AT A POINT 1.5 CM FROM THE DISTAL END. INSERT THE CHANNEL CLEANING BRUSH OR THE SINGLE USE SINGLE-ENDED CLEANING BRUSH INTO THE IRRIGATION PORT AS ILLUSTRATED BY E IN FIGURE 7.20. USING SHORT STROKES, FEED THE BRUSH THROUGH THE INSERTION SECTION UNTIL IT STOPS. IT CAN BE CONFIRMED WHETHER THE BRUSH HAS REACHED THE DISTAL END BY LOOKING AT THE TIP OF BRUSH THROUGH THE GROOVE OF THE BALLOON WATER FEEDING AND SUCTION PORT (ROUGH STANDARD OF BRUSH INSERTION LENGTH: 85 CM FROM POINT). CAREFULLY PULL THE BRUSH OUT THROUGH THE CHANNEL. CLEAN THE BRISTLES WITH YOUR FINGERTIPS IN THE DETERGENT SOLUTION. REPEAT UNTIL ALL DEBRIS IS REMOVED.
THE SERVICE CENTER WAS INFORMED DURING AN IN-SERVICE WITH AN ENDOSCOPY SUPPORT SPECIALIST (ESS) AT THE CUSTOMER¿S SITE, IT WAS NOTED THAT THE SCOPE WAS IMPROPERLY OR INCORRECTLY REPROCESSED AND USED. THE CUSTOMER INFORMED THE ESS, THE FACILITY DOES NOT USE THE RECOMMEND BW-400B SINGLE-ENDED CLEANING BRUSH TO CLEAN THE BALLOON CHANNEL OF THE SCOPE. A CHANNEL OPENING BRUSH IS USED TO CLEAN INSIDE THE CHANNEL. THERE WAS NO PATIENT INFECTION OR DEVICE POSITIVE CULTURE REPORTED. THIS IS 2 OF 2 REPORTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 546042 | EVIS EXERA II ULTRASONIC BRONCHOFIBERVIDEOSCOPE | ULTRASONIC BRONCHOFIBERVIDEOSCOPE | PSV | OLYMPUS MEDICAL SYSTEMS CORP. | BF-UC180F | 04953170356360 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |