MINIMED 780G OUS SYSTEM BLE CONNECT 3.0 MG/DL
Report
- Report Number
- 2032227-2021-134036
- Event Type
- Malfunction
- Date Received
- April 12, 2021
- Date of Event
- April 6, 2021
- Report Date
- August 17, 2021
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 000000763000384289
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER
Narratives
DEVICE PASSED ACTIVE CURRENT MEASUREMENT, SLEEP CURRENT MEASUREMENT TEST, SELFTEST, REWIND, SEATING, BASIC OCCLUSION TEST, OCCLUSION TEST, FORCE SENSOR TEST AND DISPLACEMENT TEST. DEVICE SUCCESSFULLY DOWNLOADED TRACES AND HISTORY FILE USING THUMP. DEVICE WAS CONNECTED TO A TEST METER, ACCU-CHECK GUIDE LINK, AND WAS ABLE TO CONNECT. DEVICE CONNECTED TO THE MINIMED MOBILE APP SUCCESSFULLY ON BOTH ANDROID AND IOS. THE FORMATTED HISTORY FILE CONFIRMED PUMP ERROR 53 ALARM (LINE NUMBER (B)(4) FILE NUMBER (B)(4)). PROBLEM ISOLATED TO THE ELECTRONIC ASSEMBLY AS PER GLOBAL LOGIC ANALYSIS (ESF# (B)(4)). PUMP ERROR 15 ALARM (LINE NUMBER (B)(4), FILE NUMBER (B)(4)) IS FOUND IN THE FORMATTED HISTORY FILE. PROBLEM ISOLATED TO THE ELECTRONIC ASSEMBLY AS PER GLOBAL LOGIC ANALYSIS. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: SCRATCHED CASE AND PILLOWING KEYPAD OVERLAY. THE P-CAP/RESERVOIR DOES LOCK PROPERLY. DEVICE TRANSFERRED TO RESEARCH AND DEVELOPMENT FOR FURTHER TESTING. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD THE CRITICAL BLOCK AND INVERSE CRITICAL BLOCK DID NOT ADD TO ZERO PUMP ERROR ALARM OCCURRED LAST NIGHT. CUSTOMER STATED CANNULA FILL AND AUTO TEST. THE INSULIN PUMP WAS OK. MOBILE AND METER RECONNECTED SUCCESSFULLY BUT DURING TRANSMITTER PAIRING SOFTWARE ERROR DETECTED ALARM OCCURRED AND PUMP REBOOTED. TRIED AGAIN TRANSMITTER PAIRING BUT STILL INSULIN PUMP ERROR REBOOT OCCURRED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 550040 | MINIMED 780G OUS SYSTEM BLE CONNECT 3.0 MG/DL | ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1886 | HG4R1JZZZ | 000000763000384289 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |