FDA Adverse Event Injury Summary report: N

ID NOW COVID-19 ASSAY

MDR report key: 11645215 · Received April 9, 2021

Report

Report Number
1221359-2021-00956
Event Type
Injury
Date Received
April 9, 2021
Date of Event
March 14, 2021
Report Date
July 16, 2021
Manufacturer
ABBOTT DIAGNOSTIC SCARBOROUGH, INC
Product Code
QJR
PMA / PMN Number
EUA2000074
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

H10: TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT M141753 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-000 / LOT: M141753 , TEST BASE PART NUMBER 190-430 / LOT: M141753 . THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT M141753 SHOWED THAT THE COMPLAINT RATE IS (B)(4)%. ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE HOWEVER A POSSIBLE ASSIGNABLE ROOT CAUSE IS SAMPLE INTERFERENCE OR CROSS CONTAMINATION.

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION: THIS INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE PROVIDED.

Additional Manufacturer Narrative · 1

THIS INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE PROVIDED. REFERENCE MFR REPORTS: 1221359-2021-00924, 1221359-2021-00955, 1221359-2021-00957, 1221359-2021-00958, 1221359-2021-00959, 1221359-2021-00960.

Description of Event or Problem · 1

THE CUSTOMER REPORTED FALSE POSITIVE RESULTS WITH THE ID NOW COVID-19 ASSAY FOR MULTIPLE PATIENTS PERFORMED ON MULTIPLE DATES. THIS MFR. REPORT ADDRESSES PATIENT 3 OF 7. THE CUSTOMER REPORTED A FALSE POSITIVE RESULT WITH ID NOW COVID-19 ASSAY PERFORMED ON (B)(6) 2021 ON A DIRECT TESTED NASAL KITTED SWAB. CONFIRMATION TESTING ON NASAL SWABS WITH FLUID GENERATED NEGATIVE RESULTS. FLUVID WAS PERFORMED ON (B)(6) 2021. THE CONFIRMED THE PATIENT WAS ASYMPTOMATIC. ADDITIONALLY, THE CUSTOMER CONFIRMED THAT THERE WAS A DELAY OR IMPACT IN PATIENTS TREATMENT DUE TO THE RESULTS. THE PATIENTS CARDIA CATHERIZATION WAS DELAYED. POSITIVE RESULTS ARE INDICATIVE OF SARS-COV-2 RNA; CLINICAL CORRELATION WITH PATIENT HISTORY AND OTHER DIAGNOSTIC INFORMATION IS NECESSARY TO DETERMINE PATIENT INFECTION STATUS. POSITIVE RESULTS DO NOT RULE OUT BACTERIAL INFECTION OR CO-INFECTION WITH OTHER VIRUSES. DUE TO THE HIGH SENSITIVITY OF THE ASSAYS RUN ON THE INSTRUMENT, CONTAMINATION OF THE WORK AREA WITH PREVIOUS POSITIVE SAMPLES MAY CAUSE FALSE POSITIVE RESULTS. HANDLE SAMPLES ACCORDING TO STANDARD LABORATORY PRACTICES. CLEAN INSTRUMENTS AND SURROUNDING SURFACES ACCORDING TO INSTRUCTIONS PROVIDED IN CLEANING SECTION OF THE INSTRUMENT USER MANUAL. DUE TO THE RISK OF FALSE POSITIVE RESULT LEADING TO POSSIBLE EXPOSURE TO COVID-19 THROUGH THE QUARATINE OF THE FALSE POSITIVE PATIENT WITH TRUE POSITIVE PATIENT(S), DELAYED EMERGENCY TREATMENT, UNNECESSARY TREATMENT WITH ANTIVIRAL MEDICATION, THIS EVENT SHALL BE CONSIDERED REPORTABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
542032 ID NOW COVID-19 ASSAY MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTIC SCARBOROUGH, INC M141753

Patients

Seq Age Sex Outcome Treatment
1 Other