FDA Adverse Event Injury Summary report: N

LIGHT ADJUSTABLE LENS (LAL)

MDR report key: 11645181 · Received April 9, 2021

Report

Report Number
3012712027-2021-00017
Event Type
Injury
Date Received
April 9, 2021
Date of Event
March 15, 2021
Report Date
April 9, 2021
Manufacturer
RXSIGHT, INC.
Product Code
PZK
UDI-DI
00818806020180
PMA / PMN Number
P160055
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

SITE REPORTED THAT THE LIGHT ADJUSTABLE LENS IN THE PATIENT'S LEFT EYE WAS EXPLANTED AS THE DESIRED REFRACTIVE OUTCOME WAS NOT ACHIEVED AFTER LIGHT ADJUSTMENT TREATMENTS. ANOTHER LIGHT ADJUSTABLE LENS WAS IMPLANTED IN THE EXCHANGE. THE EXPLANTED LENS WAS RETURNED FOR EVALUATION. VISUAL AND OPTICAL TESTING VERIFIED THE PRESENCE OF A ZONE NEAR THE CENTER OF THE LENS. THE PRESENCE OF THE ZONE IS INDICATIVE OF EXPOSURE OF THE LENS TO AMBIENT UV LIGHT.

Description of Event or Problem · 0

SITE REPORTED THAT THE LIGHT ADJUSTABLE LENS IN THE PATIENT'S LEFT EYE WAS EXPLANTED AS THE DESIRED REFRACTIVE OUTCOME WAS NOT ACHIEVED AFTER LIGHT ADJUSTMENT TREATMENTS. ANOTHER LIGHT ADJUSTABLE LENS WAS IMPLANTED IN THE EXCHANGE.

Additional Manufacturer Narrative · 1

SITE REPORTED THAT THE LIGHT ADJUSTABLE LENS IN THE PATIENT'S LEFT EYE WAS EXPLANTED AS THE DESIRED REFRACTIVE OUTCOME WAS NOT ACHIEVED AFTER LIGHT ADJUSTMENT TREATMENTS. ANOTHER LIGHT ADJUSTABLE LENS WAS IMPLANTED IN THE EXCHANGE. DEVICE HISTORY RECORD FOR THE LENS WAS REVIEWED. NO ISSUES WERE NOTED. THE EXPLANTED LENS HAS BEEN RECEIVED AND IS CURRENTLY UNDER EVALUATION.

Description of Event or Problem · 1

SITE REPORTED THAT THE LIGHT ADJUSTABLE LENS IN THE PATIENT'S LEFT EYE WAS EXPLANTED AS THE DESIRED REFRACTIVE OUTCOME WAS NOT ACHIEVED AFTER LIGHT ADJUSTMENT TREATMENTS. ANOTHER LIGHT ADJUSTABLE LENS WAS IMPLANTED IN THE EXCHANGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
541508 LIGHT ADJUSTABLE LENS (LAL) LIGHT ADJUSTABLE LENS (LAL) PZK RXSIGHT, INC. 60005 L02-001212 00818806020180

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention