ZILVER VENA VENOUS SELF EXPANDING STENT
Report
- Report Number
- 3001845648-2021-00292
- Event Type
- Injury
- Date Received
- April 9, 2021
- Date of Event
- October 16, 2020
- Report Date
- October 29, 2021
- Manufacturer
- COOK IRELAND LTD
- Product Code
- QAN
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
PMA/510(K) #: P200023 PRODUCT CODE: QAN DEVICE EVALUATION THE ZILVER VENA VENOUS SELF EXPANDING STENT (ZVT-7) DEVICES OF UNKNOWN LOT NUMBERS INVOLVED IN THIS COMPLAINT WERE IMPLANTED IN THE PATIENTS AND WERE NOT AVAILABLE FOR EVALUATION. WITH THE INFORMATION PROVIDED, A DOCUMENT-BASED INVESTIGATION WAS CONDUCTED. AS THE RPN AND LOT NUMBER OF THE COMPLAINT STENTS ARE UNKNOWN, A REVIEW OF THE RELEVANT MANUFACTURING RECORDS CANNOT BE CONDUCTED. HOWEVER, PRIOR TO DISTRIBUTION ALL ZILVER VENA VENOUS SELF EXPANDING STENT DEVICES ARE SUBJECT TO VISUAL A VISUAL INSPECTION AND FUNCTIONAL CHECKS TO ENSURE DEVICE INTEGRITY. THESE INSPECTIONS AND FUNCTIONAL CHECKS ARE OUTLINED IN INTERNAL PROCEDURES IN PLACE AT CIRL. THERE IS NO EVIDENCE TO SUGGEST THE USER DID NOT FOLLOW THE IFU (IFU0047-5). A DEFINITIVE ROOT CAUSE COULD NOT BE IDENTIFIED FROM THE AVAILABLE INFORMATION. A POSSIBLE ROOT CAUSE COULD BE RELATED TO PATIENTS PRE-EXISTING CONDITIONS. IT IS KNOWN THAT THAT THE PATIENTS INCLUDED IN THE STUDY HAS SUFFERED WITH EITHER ACUTE OR CHRONIC DEEP VEIN THROMBOSIS OR HAS BEEN TREATED FOR POST THROMBOTIC SYNDROME OR ILIAC VENOUS COMPRESSION SYNDROMES. IT SHOULD BE NOTED THAT RESTENOSIS IS LISTED AS A POTENTIAL ADVERSE EVENT IN THE INSTRUCTIONS FOR USE. THE COMPLAINT IS CONFIRMED BASED ON CUSTOMER TESTIMONY. ACCORDING TO THE INITIAL REPORT ALL PATIENTS WERE TREATED SUCCESSFULLY WITH THROMBOLYSIS, BALLOON ANGIOPLASTY, OR STENTING COMPLAINTS OF THIS NATURE WILL CONTINUE TO BE MONITORED FOR POTENTIAL EMERGING TRENDS.
FINAL MDR BEING SUBMITTED FUE TO COMPLETION OF THE INVESTIGATION ON 29-OCT-21
PMA/510(K) #: P200023. PRODUCT CODE: QAN. (B)(4). INVESTIGATION IS STILL PENDING, A FOLLOW UP MDR WILL BE SUBMITTED TO INCLUDE THE INVESTIGATION CONCLUSIONS.
FU, 2020, ZILVER VENA AND STENT CHARACTERISTICS OF 32 PATIENTS WITH EARLY (<14 DAYS) ILIOFEMORAL STENT OCCLUSION. PATIENTS DIAGNOSED WITH EARLY IN-STENT THROMBOSIS AFTER ILIOFEMORAL STENTING WERE REVIEWED IN THIS RETROSPECTIVE ANALYSIS. PATIENTS WITH A RECURRENCE OF SYMPTOMS OR THE DETECTION OF THROMBUS IN STENTS WITHIN 14 DAYS AFTER STENT IMPLANTATION WERE INCLUDED. PATIENTS WITH STENTS PLACED BELOW THE ORIFICE OF THE DEEP FEMORAL VEIN WERE EXCLUDED. THE PATENCY RATE WAS 90% (27/30), AND THREE CASES OF REOCCLUSION OCCURRED. ONE WAS DETECTED IN THE SECOND MONTH WITH THROMBOSIS IN THE STENT. THE OTHER TWO CASES WERE DETECTED IN MONTHS 6 AND 8 AS A RESULT OF STENOSIS OF THE STENTED SECTIONS. ALL WERE TREATED SUCCESSFULLY WITH THROMBOLYSIS, BALLOON ANGIOPLASTY, OR STENTING. THIS COMPLAINT WILL CAPTURE 2 CASES OF RESTENOSIS IN MONTHS 6 AND 8 AS A RESULT OF STENOSIS OF THE STENTED SECTIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 537697 | ZILVER VENA VENOUS SELF EXPANDING STENT | QAN | COOK IRELAND LTD |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 46 YR | Unknown | Required Intervention |