FDA Adverse Event Injury Summary report: N

ID NOW COVID-19 ASSAY

MDR report key: 11639438 · Received April 8, 2021

Report

Report Number
1221359-2021-00904
Event Type
Injury
Date Received
April 8, 2021
Report Date
February 11, 2022
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QJR
UDI-DI
10811877011320
PMA / PMN Number
EUA2000074
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H10: TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 1013553 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 191-000 / LOT: 1013553, TEST BASE PART NUMBER 190-430 / LOT: 1013553. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 1013553 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. HOWEVER A POSSIBLE ASSIGNABLE ROOT CAUSE IS SAMPLE INTERFERENCE OR CROSS CONTAMINATION. REFERENCE MFR NUMBERS: 1221359-2021-00898. 1221359-2021-00899. 1221359-2021-00900. 1221359-2021-00901. 1221359-2021-00902. 1221359-2021-00905. 1221359-2021-00928.

Additional Manufacturer Narrative · 1

REFERENCE MFR. REPORT :1221359-2021-00898, 1221359-2021-00899, 1221359-2021-00900, 1221359-2021-00901, 1221359-2021-00902, 1221359-2021-00903, AND 1221359-2021-00905. THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A TOTAL OF 27 FALSE POSITIVE RESULTS WITH THE ID NOW COVID-19 ASSAY PERFORMED (B)(6) 2021 THRU (B)(6) 2021 USING TWO DIFFERENT LOTS. THIS MFR. REPORT ADDRESSES 7 OF 8 EVENTS. A LOG FILE REVIEW ON THIS CASE IDENTIFIED TWO ADDITIONAL LOT NUMBERS 1013553 AND 1007533 AS BEING USED DURING THE SAME DATES AS LOT NUMBER 1009983. HOWEVER, NO ADDITIONAL INFORMATION REGARDING THE INVOLVEMENT OF THESE LOT NUMBERS WAS PROVIDED. POSITIVE RESULTS ARE INDICATIVE OF THE PRESENCE OF SARS-COV-2 RNA; CLINICAL CORRELATION WITH PATIENT HISTORY AND OTHER DIAGNOSTIC INFORMATION IS NECESSARY TO DETERMINE PATIENT INFECTION STATUS. POSITIVE RESULTS DO NOT RULE OUT BACTERIAL INFECTION OR CO-INFECTION WITH OTHER VIRUSES. DUE TO THE HIGH SENSITIVITY OF THE ASSAYS RUN ON THE INSTRUMENT, CONTAMINATION OF THE WORK AREA WITH PREVIOUS POSITIVE SAMPLES MAY CAUSE FALSE POSITIVE RESULTS. HANDLE SAMPLES ACCORDING TO STANDARD LABORATORY PRACTICES. CLEAN INSTRUMENTS AND SURROUNDING SURFACES ACCORDING TO INSTRUCTIONS PROVIDED IN THE CLEANING SECTION OF THE INSTRUMENT USER MANUAL. DUE TO THE RISK OF A FALSE POSITIVE RESULT LEADING TO POSSIBLE EXPOSURE TO COVID-19 THROUGH THE QUARANTINE OF THE FALSE POSITIVE PATIENT WITH TRUE POSITIVE PATIENT(S), DELAYED EMERGENCY TREATMENT AND/OR INAPPROPRIATE TREATMENT WITH ANTIVIRAL THERAPY, THE INCIDENT SHALL BE CONSIDERED REPORTABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
530013 ID NOW COVID-19 ASSAY MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 1013553 10811877011320

Patients

Seq Age Sex Outcome Treatment
1 Unknown Hospitalization