FDA Adverse Event Malfunction Summary report: N

GAMMA-CLONE ANTI-C (MONOCLONAL) SERIES 2

MDR report key: 11639337 · Received April 8, 2021

Report

Report Number
1034569-2021-00009
Event Type
Malfunction
Date Received
April 8, 2021
Date of Event
February 8, 2021
Report Date
April 8, 2021
Manufacturer
IMMUCOR, INC.
Product Code
QHR
UDI-DI
10888234001355
PMA / PMN Number
103498/5033
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ON FEBRUARY 9, 2021 IMMUCOR CONFIRMED REACTIVITY OF RETENTION ANTI-C MONOCLONAL LOT 936210A WITH FOUR HETEROZYGOUS C ANTIGEN POSITIVE CELLS AND FOUR C ANTIGEN NEGATIVE CELLS USING RETENTION PANOCELL-20 LOT 50556 AND PANOCELL-16 LOT 02592. CONTROLS PERFORMED AS EXPECTED AND ALL THE C POSITIVE CELLS RESULTED AS EXPECTED. TESTING PERFORMED PER PACKAGE INSERT INSTRUCTIONS. RETENTION PRODUCT PERFORMED AS EXPECTED. ON FEBRUARY 11, 2021 IMMUCOR CONFIRMED REACTIVITY OF RETURN ANTI-C MONOCLONAL LOT 936210A WITH FOUR HETEROZYGOUS C ANTIGEN POSITIVE CELLS AND FOUR C ANTIGEN NEGATIVE CELLS USING RETENTION PANOCELL-20 LOT 50556 AND PANOCELL-16 LOT 02592. CONTROLS PERFORMED AS EXPECTED AND ALL THE C POSITIVE CELLS RESULTED AS EXPECTED. TESTING PERFORMED PER PACKAGE INSERT INSTRUCTIONS. RETURN PRODUCT PERFORMED AS EXPECTED. ON FEBRUARY 11, 2021 IMMUCOR TESTED RETURN SEGMENTS FROM UNIT W0352 21 11693600W WITH RETENTION AND RETURN ANTI-C (MONOCLONAL) GAMMA-CLONE LOT 936210A. CONTROLS PERFORMED AS EXPECTED. THE SEGMENT FROM UNIT W0352 21 11693600W TESTED NEGATIVE WITH BOTH RETENTION AND RETURN ANTI-C (MONOCLONAL) GAMMA-CLONE LOT 936210A. THE IMMUCOR INTERNAL RECORD NUMBER FOR THIS REPORT IS (B)(4).

Description of Event or Problem · 1

ON MARCH 10, 2021 IMMUCOR BECAME AWARE OF AN EVENT INVOLVING AN UNEXPECTED POSITIVE REACTION FOR A UNIT FROM A DONOR WHO WAS HISTORICALLY NEGATIVE FOR C ANTIGEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
530005 GAMMA-CLONE ANTI-C (MONOCLONAL) SERIES 2 ANTI-C (MONOCLONAL) BLOOD GROUPING REAGENT QHR IMMUCOR, INC. 936210-A 10888234001355

Patients

Seq Age Sex Outcome Treatment
1