FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 1163380 · Received September 10, 2008

Report

Report Number
1824206-2008-03373
Event Type
Malfunction
Date Received
September 10, 2008
Date of Event
August 1, 2008
Report Date
August 1, 2008
Manufacturer
HILL-ROM, RITTER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

CUSTOMER STATED THAT NONE OF THE HYDRAULIC FUNCTIONS WERE WORKING. FOUND THE HYDRAULIC RESERVOIR TANK WAS EMPTY OF ALL FLUID. AND, THERE WAS EVIDENCE OF LEAKAGE FROM MANIFOLD. REPLACED HYDRAULIC MANIFOLD TO RESOLVE ISSUE, NO INJURY BEING REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC POWERED HOSPITAL BED FNL HILL-ROM, RITTER 1900 P1900AF1619

Patients

Seq Age Sex Outcome Treatment
1