FDA Adverse Event Malfunction Summary report: N

VERSACARE

MDR report key: 1163376 · Received September 10, 2008

Report

Report Number
1824206-2008-03368
Event Type
Malfunction
Date Received
September 10, 2008
Date of Event
August 14, 2008
Report Date
August 14, 2008
Manufacturer
HILL-ROM RITTER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

A HILL-ROM TECHNICAL SERVICE REP FOUND THAT THE GROUND RESISTANCE READING WAS HIGH WHILE PERFORMING PREVENTATIVE MAINTENANCE AND THERE WERE NO REPORTED INJURIES. HE REPLACED THE POWER CORD AND THIS RESOLVED THE LOSS OF GROUND.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VERSACARE AC POWERED HOSPITAL BED FNL HILL-ROM RITTER P3200 NA

Patients

Seq Age Sex Outcome Treatment
1