FDA Adverse Event
Malfunction
Summary report: N
VERSACARE
MDR report key: 1163376
·
Received September 10, 2008
Report
- Report Number
- 1824206-2008-03368
- Event Type
- Malfunction
- Date Received
- September 10, 2008
- Date of Event
- August 14, 2008
- Report Date
- August 14, 2008
- Manufacturer
- HILL-ROM RITTER
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
A HILL-ROM TECHNICAL SERVICE REP FOUND THAT THE GROUND RESISTANCE READING WAS HIGH WHILE PERFORMING PREVENTATIVE MAINTENANCE AND THERE WERE NO REPORTED INJURIES. HE REPLACED THE POWER CORD AND THIS RESOLVED THE LOSS OF GROUND.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VERSACARE | AC POWERED HOSPITAL BED | FNL | HILL-ROM RITTER | P3200 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |