Summary report: N
M SERIES
MDR report key: 1163366
·
Received September 9, 2008
Report
- Report Number
- 1220908-2008-02099
- Event Type
- Malfunction
- Date Received
- September 9, 2008
- Report Date
- August 19, 2008
- Manufacturer
- ZOLL MEDICAL CORPORATION
- Product Code
- MKJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
ZOLL MEDICAL CORPORATION HAS RECEIVED THE PRODUCT AND WILL BE PROVIDING A FOLLOW-UP REPORT WHEN THE INVESTIGATION IS COMPLETED.
Description of Event or Problem · 1
COMPLAINANT ALLEGED THAT DURING BIOMED TESTING THE DEVICE'S PACER OUTPUT WAS OBTAINING INCONSISTENT READINGS. WHEN SET AT 60 MA, THE DEVICE WAS READING AT 65 TO 68 MA. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | M SERIES | DEFIBRILLATOR/PACEMAKER | MKJ | ZOLL MEDICAL CORPORATION | MSERIES BI-PHASIC | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |