FDA Adverse Event Malfunction Summary report: N

ULTRACISION DISSECT HOOK 10CM

MDR report key: 1163336 · Received September 10, 2008

Report

Report Number
3005075853-2008-01625
Event Type
Malfunction
Date Received
September 10, 2008
Date of Event
July 2, 2008
Report Date
July 22, 2008
Manufacturer
ETHICON ENDO-SURGERY, LLC
Product Code
LFL
PMA / PMN Number
K012176
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVAL SUMMARY: THE DEVICES WERE RETURNED WITH THE BLADE SCRATCHED AND CRACKED. DURING FUNCTIONAL TESTING, AN ERROR CODE 5 WAS DISPLAYED AND THE BLADE TIP FURTHER CRACKED ON BOTH DEVICES. THE DEVICE WILL STOP ACTIVATING, AND EITHER EMIT A SOLID TONE OR DISPLAY AN INSTRUMENT ERROR CODE 5 OR A SOLID TONE ON THE GENERATOR DISPLAY WHEN THE BLADE BECOMES DAMAGED. WHEN A BLADE HAS BEEN COMPROMISED SUCH AS SCRATCHED OR CRACKED, THE TITANIUM METAL IS FATIGUED WHEN CONTINUALLY ENERGIZED AND THIS RESULTS IN THE BLADE FURTHER CRACKING AND POSSIBLY BREAKING OFF. A POSSIBLE CAUSE OF AN ERROR CODE 5 (INSTRUMENT ERROR) IS BLADE DAMAGE. BLADE DAMAGE MAY OCCUR FROM EXTERNAL CONTACT WITH METAL OR PLASTIC DURING PRE-OP OR GENERAL USE. EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED PRIOR TO SHIPMENT, AND DAMAGE OF THIS MAGNITUDE WOULD HAVE BEEN DETECTED AT THIS PROCESS. COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MFG PROCESS. SECOND DEVICE BATCH # E9EV2Y. MFG DATE: 03/2008. EXP DATE: 02/2013.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A CABG PROCEDURE, THE DEVICES LOCKED OUT. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PT.