HS LONG 5MM CURVED SHEARS
Report
- Report Number
- 3005075853-2008-01627
- Event Type
- Malfunction
- Date Received
- September 10, 2008
- Date of Event
- March 6, 2008
- Report Date
- August 12, 2008
- Manufacturer
- ETHICON ENDO-SURGERY, LLC
- Product Code
- GEI
- PMA / PMN Number
- K002981
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- EC
- Reporter Occupation
- OTHER
Narratives
EVALUATION SUMMARY - THE DEVICE WAS RETURNED WITH THE DISTAL TIP OF THE BLADE BROKEN OFF AND MISSING. THE REMAINING BLADE PORTION WAS SCRATCHED. THIS BLADE TIP PORTION MAY HAVE BROKEN OFF THE DEVICE DURING TRANSPORT TO OUR ANALYSIS SITE. THE REPORTED COMPLAINT WAS FOR AN ACTIVATION ISSUE. A POSSIBLE CAUSE OF AN ACTIVATION ISSUE IS BLADE DAMAGE. BLADE DAMAGE IS CAUSED BY CONTACTING AN ACTIVE BLADE WITH OTHER SURGICAL DEVICES, STAPLES OR CLIPS DURING THE PROCEDURE OR USING ANY MEANS OTHER THAN THE BLADE WRENCH TO ATTACH OR DETACH THE BLADE. ONCE BLADE DAMAGE HAS OCCURRED, SUBSEQUENT ACTIVATIONS MAY RESULT IN FATIGUE FAILURE INITIATING AT THE ORIGINAL DAMAGE LOCATION, WHICH CAN RESULT IN THE BLADE TIP BREAKING OFF AS DESCRIBED ABOVE. EACH DEVICE IS VISUALLY INSPECTED AN FUNCTIONALLY TESTED PRIOR TO SHIPMENT, AND DAMGE OF THIS MAGNITUDE WOULD HAVE BEEN DETECTED AT THIS PROCESS. COMPLAINT INFORMATION IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE DEVICE STOPPED WORKING AFTER 6 HOURS. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HS LONG 5MM CURVED SHEARS | GEI | ETHICON ENDO-SURGERY, LLC | NA | D4GV51 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR| HANDPIECE |