FDA Adverse Event Malfunction Summary report: N

PUMP MMT-1780KPK 670G PATHWAY BLACK MG

MDR report key: 11632423 · Received April 8, 2021

Report

Report Number
2032227-2021-132870
Event Type
Malfunction
Date Received
April 8, 2021
Date of Event
April 4, 2021
Report Date
April 26, 2022
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
UDI-DI
000000763000090203
Removal / Correction Number
Z-0956-2020
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS PART OF A RETROSPECTIVE REVIEW AND REMEDIATION EFFORTS IN RESPONSE TO A WARNING LETTER. UPDATED H9: Z-0956-2020. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Additional Manufacturer Narrative · 1

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

DEVICE P-CAP OR RESERVOIR LOCKED PROPERLY IN PLACE. DEVICE RECEIVED WITH CRACKED RETAINER, LABEL DAMAGE, CRACKED CASE (BATTERY TUBE), CRACKED CASE-CORNER OF BELT CLIP RAILS, AND CRACKED BATTERY TUBE THREADS. DEVICE RECEIVED WITH CRITICAL PUMP ERROR DUE TO MOISTURE DAMAGE TO FORCE SENSOR. DEVICE RECEIVED WITH CORRODED ELECTRONIC ASSEMBLIES AND CORRODED MOTOR HOME SWITCH. DEVICE UNABLE TO PERFORM DISPLACEMENT TEST, REWIND, PRIME OR SEATING, BASIC OCCLUSION, FORCE SENSOR, OCCLUSION, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT, AND SELF TEST DUE TO CRITICAL PUMP ERROR (OPEN BOOK IMAGE) ALARM.

Description of Event or Problem · 1

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD CRITICAL PUMP ERROR ALARM AND OPEN BOOK IMAGE WAS DISPLAYED ON THE SCREEN. CUSTOMER WAS HIKING AND FELL INTO RIVER. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
530742 PUMP MMT-1780KPK 670G PATHWAY BLACK MG ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL OZP MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1780KPK HG2HB16 000000763000090203

Patients

Seq Age Sex Outcome Treatment
1 45 YR Male