FDA Adverse Event Malfunction Summary report: N

BABY GORILLA/GORILLA PLATING SYSTEM

MDR report key: 11628898 · Received April 7, 2021

Report

Report Number
3008650117-2021-00080
Event Type
Malfunction
Date Received
April 7, 2021
Date of Event
March 2, 2021
Report Date
April 7, 2021
Manufacturer
PARAGON 28, INC.
Product Code
HWC
PMA / PMN Number
K190365
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE HISTORY RECORD FOR LOT NUMBER TC29409R WAS REVIEWED AND SHOWS ALL INSPECTED PARTS WERE RECEIVED AND ACCEPTED. NO NCR IDENTIFIED ON DHR. ON 4/19/2021, THE PART WAS RECEIVED. VISUAL OBSERVATION OF THE RETURED PART CONFIRMED THE FAILURE MODE; DRIVER BROKEN AT THE TIP. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Additional Manufacturer Narrative · 1

THE DEVICE HISTORY RECORD FOR LOT NUMBER TC29409R WAS REVIEWED AND SHOWS ALL INSPECTED PARTS WERE RECEIVED AND ACCEPTED. NO NCR IDENTIFIED ON DHR. AS OF 4/7/2021, THE PART WAS NOT RETURNED TO THE MANUFACTURER. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE ON (B)(6) 2021, A PARAGON 28 SCREW DRIVER BROKE WHILE EXPLANTING A SCREW THAT WAS TOO LONG. THE REPORTER STATED THAT THE BREAKAGE DID NOT INTERFERE WITH THE IMPLANT PLACEMENT. IT WAS REPORTED THAT THE SURGERY WAS DELAYED FOR MORE THAN 30 MINUTES, IN THE ATTEMPT TO FIND A SUITABLE DRIVER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
529125 BABY GORILLA/GORILLA PLATING SYSTEM SCREW DRIVER ATTACHMENT, SOLID, LONG, AO, HX-15 HWC PARAGON 28, INC. P99-191-SB15 TC29409R

Patients

Seq Age Sex Outcome Treatment
1 74 YR P99-000-AOMN MINI-AO, RATCHETING HANDLE