FDA Adverse Event Malfunction Summary report: N

HAND PIECE FOR BATTERY POWERED DRIVER

MDR report key: 11628495 · Received April 7, 2021

Report

Report Number
2939274-2021-01722
Event Type
Malfunction
Date Received
April 7, 2021
Report Date
March 17, 2021
Manufacturer
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
Product Code
HWE
UDI-DI
10887587024585
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE RUNS LOW IN FAST FORWARD AND FORWARD AND RUNS INTERMITTENTLY IN REVERSE. DAMAGED COMPONENT IS THE REASON FOR REPAIR. THE CAUSE OF THE ISSUE IS DAMAGED COMPONENT. THE FOLLOWING PARTS WERE REPLACED: CIRCUIT BOARD, MOTOR, MEMBRANE SWITCH/FLEX CIRCUIT, AND ALL APPLICABLE COMPONENTS. THE ITEM WAS REPAIRED PER THE INSPECTION SHEET, PASSED SYNTHES FINAL INSPECTION ON (B)(6) 2021 AND WILL BE RETURNED TO THE CUSTOMER UPON COMPLETION OF THE SERVICE AND REPAIR PROCESS. FINALIZED SERVICE RECORD WILL BE ARCHIVED IN TUNGSTEN DOCUMENT MANAGEMENT SYSTEM. THE EVALUATION WAS CONFIRMED. THE DEVICE WAS DEEMED SERVICEABLE AND WILL BE RETURNED TO THE CUSTOMER, NO DESIGN OR MANUFACTURING ISSUES WERE IDENTIFIED THEREFORE IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. DEVICE HISTORY LOT PART NUMBER:05.000.008. SYNTHES LOT NUMBER: 5882258. SUPPLIER LOT NUMBER: N/A. RELEASE TO WAREHOUSE DATE: 08OCT2008. EXPIRATION DATE: N/A. SUPPLIER: (B)(4). NO NCRS WERE GENERATED DURING PRODUCTION. DEVICE HISTORY BATCH: NULL. REVIEW OF THE DEVICE HISTORY RECORD SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THIS PRODUCT, AND ANY SUBCOMPONENTS, WHICH WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT ON (B)(6), 2021, DURING A WEEKLY AUDIT IT WAS FOUND OUT THAT THE HAND PIECE FOR BATTERY POWERED DRIVER REVERSE WORKS ON/OFF. THERE WAS NO PATIENT INVOLVEMENT. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS FOR ONE (1) HAND PIECE FOR BATTERY POWERED DRIVER. THIS IS REPORT 1 OF 1 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
525668 HAND PIECE FOR BATTERY POWERED DRIVER INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT HWE WRIGHTS LANE SYNTHES USA PRODUCTS LLC 05.000.008 5882258 10887587024585

Patients

Seq Age Sex Outcome Treatment
1