FDA Adverse Event Injury Summary report: N

DRAGER (FORMALLY AIR SHIELDS)

MDR report key: 11628028 · Received April 7, 2021

Report

Report Number
MW5100584
Event Type
Injury
Date Received
April 7, 2021
Date of Event
March 29, 2021
Report Date
April 6, 2021
Manufacturer
DRAEGER MEDICAL SYSTEMS, INC.
Product Code
FMT
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
PA, US
Reporter Occupation
501

Narratives

Description of Event or Problem · 1

EVENING OF (B)(6) 2021 THERE WAS AN NEWBORN PLACED INTO THE INFANT WARMER IN THE NICU FOR EXAMINATION AND A X-RAY. THE WARMING HEAD WAS TURNED TO THE SIDE TO ALLOW THE X-RAY UNIT ACCESS TO THE INFANT. THE INFANT WAS IMMEDIATELY REMOVED FROM THE WARMER AND PLACED BACK INTO ANOTHER WARMING UNIT, AND TAKEN BACK TO THE MOTHERS ROOM. THE INFANT WARMER IN THE NICU WAS LEFT TURNED ON AND IN SKIN SENSING MODE, WHICH CAUSED THE ELEMENT TO RAMP UP TO 100% POWER OUTPUT. THE WARMING HEAD WAS ALSO LEFT IN THE TURNED POSITION DIRECTLY OVER TOP OF AN OXYGEN BLENDER. APPROX 4 HOURS LATER THE OXYGEN BLENDER MELTED DOWN AND BURST INTO FLAMES. THE PRESSURE FROM THE O2 ESCAPING FROM THE DAMAGED BLENDER SHOT BURNING PLASTIC AND MOLTEN ALUMINUM AT THE WALL BEHIND BOTH UNITS. THE ITEMS THAT WERE ON THE WALL THEN ALSO IGNITED. THE FIRE ALARM DID SOUND. HOSPITAL STAFF DID RESPOND AND ACT APPROPRIATELY, EXTINGUISHING THE FIRE WITH 3 - 10 LB ABC EXTINGUISHERS. THE ENTIRE OB DEPARTMENT, BOTH MOMS AND BABIES, WERE EVACUATED TO OTHERS PARTS OF THE BUILDING. THE FIRE DEPARTMENT RESPONDED, THEY CHECKED FOR FIRE EXTENSION AND VENTILATED SMOKE FROM THE AFFECTED AREAS. THE (B)(6) STATE POLICE FIRE MARSHAL AND THE (B)(6) DOH WERE BOTH NOTIFIED. FDA SAFETY REPORT ID# (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
527782 DRAGER (FORMALLY AIR SHIELDS) WARMER, INFANT RADIANT FMT DRAEGER MEDICAL SYSTEMS, INC. IICS-90

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention