FDA Adverse Event Malfunction Summary report: N

TOTALCARE

MDR report key: 1162754 · Received September 8, 2008

Report

Report Number
1824206-2008-03330
Event Type
Malfunction
Date Received
September 8, 2008
Date of Event
July 15, 2008
Report Date
July 15, 2008
Manufacturer
HILL-ROM RITTER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

CUSTOMER STATED THAT THE HYDRAULICS ON BED WASN'T WORKING. TSR FOUND THAT THE HEAD HI/LO UP AND THE HEAD UP FUNCTIONS OF THE MANIFOLD WEREN'T OPERATING AT ALL. TSR FOUND WEAK MANIFOLD PRESSURE. TSR REPLACED MANIFOLD. BED IS NOW REPAIRED AND PLACED INTO SERVICE. NO INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE AC POWERED HOSPITAL BED FNL HILL-ROM RITTER 1900 P1900D004031

Patients

Seq Age Sex Outcome Treatment
1