FDA Adverse Event Injury Summary report: N

STEM: AMISTEM H HA COATED STD SIZE 4

MDR report key: 11626185 · Received April 7, 2021

Report

Report Number
3005180920-2021-00267
Event Type
Injury
Date Received
April 7, 2021
Date of Event
March 8, 2021
Report Date
April 7, 2021
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030804106
PMA / PMN Number
K093944
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 06 APRIL 2021: LOT 133211: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 24-OCT-2013. EXPIRATION DATE: 2018-09-30. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, ALL ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY OTHER SIMILAR REPORTED EVENT SINCE 2017.

Description of Event or Problem · 1

THE PATIENT CAME IN REPORTING PAIN DUE TO A LOOSE STEM AND THE CAUSE OF THE LOOSE STEM IS UNKNOWN. THE SURGEON REVISED THE STEM AND HEAD WITH COMPETITOR IMPLANTS AND REVISED THE MEDACTA LINER WITH A MEDACTA LINER ALMOST 6 YEARS AND 11 MONTHS AFTER PRIMARY. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
528035 STEM: AMISTEM H HA COATED STD SIZE 4 FEMORAL CEMENTLESS STEM LZO MEDACTA INTERNATIONAL SA 01.18.134 133211 07630030804106

Patients

Seq Age Sex Outcome Treatment
1 62 YR Required Intervention