FDA Adverse Event Malfunction Summary report: N

HIP GENERAL OFFSET-30 BROACH HANDLE - LEFT

MDR report key: 11626168 · Received April 7, 2021

Report

Report Number
3005180920-2021-00269
Event Type
Malfunction
Date Received
April 7, 2021
Date of Event
March 9, 2021
Report Date
April 7, 2021
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LXH
UDI-DI
07630040742122
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 18 MARCH 2021: LOT 1111998: 11 ITEMS MANUFACTURED AND RELEASED ON 02-FEB-2012. NO ANOMALIES FOUND RELATED TO THE PROBLEM. SINCE 2017 TO DATE, NO OTHER SIMILAR EVENTS HAVE BEEN REPORTED ON THIS LOT. ADDITIONAL DEVICE INVOLVED: HIP GENERAL 01.10.10.131 OFFSET-30° BROACH HANDLE - LEFT LOT. 1111157. BATCH REVIEW PERFORMED ON 18 MARCH 2021: LOT 1111157: 15 ITEMS MANUFACTURED AND RELEASED ON 12-SEPT-2011. NO ANOMALIES FOUND RELATED TO THE PROBLEM. SINCE 2017 TO DATE, NO OTHER SIMILAR EVENTS HAVE BEEN REPORTED ON THIS LOT. ADDITIONAL DEVICE INVOLVED: HIP GENERAL 01.15.10.0133 STRAIGHT AMIS BROACH HANDLE LOT. 1952432 BATCH REVIEW PERFORMED ON 18 MARCH 2021: LOT 1952432: 60 ITEMS MANUFACTURED AND RELEASED ON 03-SEPT-19. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, ANOTHER SIMILAR EVENT HAS BEEN REPORTED ON THIS LOT.

Description of Event or Problem · 1

DURING THE SAME SURGERY THE BROACHES HANDLES DISASSEMBLED WHILE THE SURGEON WAS SWITCHING FROM A BROACH TO THE NEXT ONE. THE SURGERY WAS COMPLETED SUCCESSFULLY USING EXTRA STRAIGHT BROACH HANDLES, BUT THERE WERE 40 MINUTES OF DELAY DUE TO THIS ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
528965 HIP GENERAL OFFSET-30 BROACH HANDLE - LEFT SURGICAL INSTRUMENT FOR HIP SURGERY LXH MEDACTA INTERNATIONAL SA 01.10.10.131 1111998 07630040742122

Patients

Seq Age Sex Outcome Treatment
1 55 YR Other