FDA Adverse Event
Other
Summary report: N
PRC 5501 DE
MDR report key: 116261
·
Received August 28, 1997
Report
- Report Number
- 2027060-1997-00006
- Event Type
- Other
- Date Received
- August 28, 1997
- Date of Event
- August 23, 1997
- Report Date
- August 28, 1997
- Manufacturer
- ALM SA
- Product Code
- FSY
- Adverse Event
- Yes
- Report Source
- Distributor report
- Reporter Location
- GA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Description of Event or Problem · 1
ON 08/25/97 IT WAS REPORTED TO ALM THAT AN ALM LIGHT FIXTURE, MODEL PRC 5501 HAD DROPPED FROM THE CEILING MOUNT ON 08/23/97. ALM IMMEDIATELY DISPATCHED AN ALM TECHNICAN, ALONG WITH A REPLACEMENT LIGHT FIXTURE TO: 1) EVALUATE THE FIXTURE IN QUESTION 2) REPLACE THE FIXTURE SO THE CUSTOMER COULD CONTINUE TO WORK IN THE ROOM. THE LIGHT FIXTURE WAS REPLACED ON 08/26/97 AND THE INITIAL INSPECTION HAS SHOWN THAT SIX NON-REMOVABLE, MACHINE PRESSED PINS HAD SHEARED OFF, CAUSING THE LIGHT TO DETACH FROM THE MOUNT. NO ADDITIONAL INFORMATION CAN BE GIVEN AT THIS TIME AS THE HOSPITAL WILL NOT RELEASE THE LIGHT FIXTURE FOR FURTHER INSPECTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PRC 5501 DE | SURGICAL LIGHT | FSY | ALM SA | 555012 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | Other |