FDA Adverse Event Injury Summary report: N

CHLORAPREP ONE STEP

MDR report key: 11625637 · Received April 6, 2021

Report

Report Number
3004932373-2021-00172
Event Type
Injury
Date Received
April 6, 2021
Date of Event
March 25, 2021
Report Date
September 14, 2021
Manufacturer
CAREFUSION 213, LLC 0113
Product Code
KXF
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

PR 2659127 FOLLOW UP EMDR SUBMISSION. FAILURE MODE IS NOT CONFIRMED SINCE DEFECTIVE SAMPLE WAS NOT DELIVERED AND NO PHOTOS WERE PROVIDED BY CUSTOMER. PRODUCTION BATCH HISTORY RECORDS FOR APPLICATOR MATERIAL NO.: 930599NSB BATCH NO.: 0160547 WAS REVIEWED AND NO NON-CONFORMITIES, FAILURES, DEVIATIONS, OR REWORK ACTIVITIES OCCURRED DURING THE MANUFACTURING OF THIS LOT SIMILAR TO THE REPORTED ISSUE OR THAT COULD HAVE CONTRIBUTED TO THE REPORTED FAILURE MODE. HOWEVER SOME OPPORTUNITIES WERE FOUND IN THE CORRECTIVE MAINTENANCE OF THE FILLING AMPOULE EQUIPMENT WHICH POTENTIALLY CAN PRODUCE BROKEN AMPOULE. CORRECTIVE MAINTENANCE WERE PERFORMED TO ELIMINATE/DECREASE BROKEN AMPULE FAILURE MODE. AWARENESS TRAINING REGARDING BROKEN AMPULE WAS PROVIDED TO PRODUCTION, ENGINEERING AND QUALITY ASSOCIATES. SEE NARRATIVE BELOW.

Description of Event or Problem · 0

MATERIAL NO.: 930599NSB BATCH NO.: 0160547 IT WAS REPORTED THAT GLASS SHARDS CAME OUT OF THE CHLORAPREP APPLICATOR AND CUT A PATIENT. PER SCAR: RECEIVED CUSTOMER COMPLAINT THAT GLASS SHARD WERE COMING OUT OF ONE OF THE CHLORAPREP APPLICATORS AND CUT A PATIENT. CUSTOMER IS RETURNING THE SAMPLE IF NEEDED FOR EVALUATION.

Additional Manufacturer Narrative · 1

(B)(6) INITIAL EMDR SUBMISSION. A FOLLOW UP EMDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE. ANNEX E- E1701 ABRASION. ANNEX F- F25 UNANTICIPATED ADVERSE DEVICE EFFECT. ANNEX A- A24- ADVERSE EVENT WITHOUT IDENTIFIED DEVICE OR PROBLEM USE.

Description of Event or Problem · 1

MATERIAL NO.: 930599NSB, BATCH NO.: 0160547. IT WAS REPORTED THAT GLASS SHARDS CAME OUT OF THE CHLORAPREP APPLICATOR AND CUT A PATIENT. PER SCAR: RECEIVED CUSTOMER COMPLAINT THAT GLASS SHARD WERE COMING OUT OF ONE OF THE CHLORAPREP APPLICATORS AND CUT A PATIENT. CUSTOMER IS RETURNING THE SAMPLE IF NEEDED FOR EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
514758 CHLORAPREP ONE STEP 2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL KXF CAREFUSION 213, LLC 0113 0160547

Patients

Seq Age Sex Outcome Treatment
1 62 YR Other