CHLORAPREP ONE STEP
Report
- Report Number
- 3004932373-2021-00172
- Event Type
- Injury
- Date Received
- April 6, 2021
- Date of Event
- March 25, 2021
- Report Date
- September 14, 2021
- Manufacturer
- CAREFUSION 213, LLC 0113
- Product Code
- KXF
- PMA / PMN Number
- EXEMPT
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- 003
Narratives
PR 2659127 FOLLOW UP EMDR SUBMISSION. FAILURE MODE IS NOT CONFIRMED SINCE DEFECTIVE SAMPLE WAS NOT DELIVERED AND NO PHOTOS WERE PROVIDED BY CUSTOMER. PRODUCTION BATCH HISTORY RECORDS FOR APPLICATOR MATERIAL NO.: 930599NSB BATCH NO.: 0160547 WAS REVIEWED AND NO NON-CONFORMITIES, FAILURES, DEVIATIONS, OR REWORK ACTIVITIES OCCURRED DURING THE MANUFACTURING OF THIS LOT SIMILAR TO THE REPORTED ISSUE OR THAT COULD HAVE CONTRIBUTED TO THE REPORTED FAILURE MODE. HOWEVER SOME OPPORTUNITIES WERE FOUND IN THE CORRECTIVE MAINTENANCE OF THE FILLING AMPOULE EQUIPMENT WHICH POTENTIALLY CAN PRODUCE BROKEN AMPOULE. CORRECTIVE MAINTENANCE WERE PERFORMED TO ELIMINATE/DECREASE BROKEN AMPULE FAILURE MODE. AWARENESS TRAINING REGARDING BROKEN AMPULE WAS PROVIDED TO PRODUCTION, ENGINEERING AND QUALITY ASSOCIATES. SEE NARRATIVE BELOW.
MATERIAL NO.: 930599NSB BATCH NO.: 0160547 IT WAS REPORTED THAT GLASS SHARDS CAME OUT OF THE CHLORAPREP APPLICATOR AND CUT A PATIENT. PER SCAR: RECEIVED CUSTOMER COMPLAINT THAT GLASS SHARD WERE COMING OUT OF ONE OF THE CHLORAPREP APPLICATORS AND CUT A PATIENT. CUSTOMER IS RETURNING THE SAMPLE IF NEEDED FOR EVALUATION.
(B)(6) INITIAL EMDR SUBMISSION. A FOLLOW UP EMDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION BECOMES AVAILABLE. ANNEX E- E1701 ABRASION. ANNEX F- F25 UNANTICIPATED ADVERSE DEVICE EFFECT. ANNEX A- A24- ADVERSE EVENT WITHOUT IDENTIFIED DEVICE OR PROBLEM USE.
MATERIAL NO.: 930599NSB, BATCH NO.: 0160547. IT WAS REPORTED THAT GLASS SHARDS CAME OUT OF THE CHLORAPREP APPLICATOR AND CUT A PATIENT. PER SCAR: RECEIVED CUSTOMER COMPLAINT THAT GLASS SHARD WERE COMING OUT OF ONE OF THE CHLORAPREP APPLICATORS AND CUT A PATIENT. CUSTOMER IS RETURNING THE SAMPLE IF NEEDED FOR EVALUATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 514758 | CHLORAPREP ONE STEP | 2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL | KXF | CAREFUSION 213, LLC 0113 | 0160547 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Other |