ID NOW COVID-19 ASSAY
Report
- Report Number
- 1221359-2021-00420
- Event Type
- Malfunction
- Date Received
- April 6, 2021
- Date of Event
- June 12, 2020
- Report Date
- April 6, 2021
- Manufacturer
- ABBOTT DIAGNOSTICS SCARBOROUGH INC.
- Product Code
- QJR
- UDI-DI
- 10811877011269
- PMA / PMN Number
- EUA2000074
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS (B)(4), INC. ON RETAINED KIT LOT M120401 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 190-000 / LOT: M120401, TEST BASE PART NUMBER 190-430 / LOT: M120401 THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW COMPLAINTS REPORTED AS FALSE POSITIVE RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT M120401 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS (B)(4), INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. THE AVAILABLE EVIDENCE SUGGESTS THAT THIS DEVICE LOT IS PERFORMING WITHIN THE STATEMENTS MADE WITHIN PACKAGE INSERT RELATED TO % TEST AGREEMENT WITH THE EXPECTED RESULTS BY SAMPLE CONCENTRATION. REFERENCE MANUFACTURE REPORT NUMBERS: 1221359-2021-00421, 221359-2021-00422, AND 1221359-2021-00423.
THE CUSTOMER REPORTED FALSE POSITIVE WITH THE ID NOW COVID-19 ASSAY FOR MULTIPLE PATIENTS. THERE WERE FOUR (4) FALSE POSITIVE RESULTS REPORTED. THIS MFR REPORT REPRESENT PATIENT ONE ( 1) OF FOUR (4), TEST PERFORMED ON (B)(6) 2020 WITH LOT M120401. THE CUSTOMER REPORTED THAT THE ID NOW COVID-19 ASSAY WAS PERFORMED ON A DIRECT TESTED NASAL SWAB. THE NASAL SWAB WAS USED TO SWAB BOTH NOSTRILS PER THE PRODUCT INSERT INSTRUCTIONS. REPEAT TESTING WAS NOT PERFORMED. PCR CONFIRMATION TESTING WAS PERFORMED ON A NASOPHARYNGEAL SWAB IN VIRAL TRANSPORT MEDIA WITH BDMAX GENERATED NEGATIVE RESULT. CT VALUE NOT PROVIDED. THE CUSTOMER STATED THE PATIENT WAS ASYMPTOMATIC. NO ADDITIONAL PATIENT INFORMATION, INCLUDING TREATMENT AND OUTCOME, WAS PROVIDED. POSITIVE RESULTS ARE INDICATIVE OF THE PRESENCE OF SARS-COV-2 RNA; CLINICAL CORRELATION WITH PATIENT HISTORY AND OTHER DIAGNOSTIC INFORMATION IS NECESSARY TO DETERMINE PATIENT INFECTION STATUS. POSITIVE RESULTS DO NOT RULE OUT BACTERIAL INFECTION OR CO-INFECTION WITH OTHER VIRUSES. DUE TO THE HIGH SENSITIVITY OF THE ASSAYS RUN ON THE INSTRUMENT, CONTAMINATION OF THE WORK AREA WITH PREVIOUS POSITIVE SAMPLES MAY CAUSE FALSE POSITIVE RESULTS. HANDLE SAMPLES ACCORDING TO STANDARD LABORATORY PRACTICES. CLEAN INSTRUMENTS AND SURROUNDING SURFACES ACCORDING TO INSTRUCTIONS PROVIDED IN THE CLEANING SECTION OF THE INSTRUMENT USER MANUAL. DUE TO THE RISK OF A FALSE POSITIVE RESULT LEADING TO POSSIBLE EXPOSURE TO COVID-19 THROUGH THE QUARANTINE OF THE FALSE POSITIVE PATIENT WITH TRUE POSITIVE PATIENT(S), DELAYED EMERGENCY TREATMENT AND/OR INAPPROPRIATE TREATMENT WITH ANTIVIRAL THERAPY, THE INCIDENT SHALL BE CONSIDERED REPORTABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 521220 | ID NOW COVID-19 ASSAY | MOLECULAR IVD FOR ID NOW COVID-19 | QJR | ABBOTT DIAGNOSTICS SCARBOROUGH INC. | M120401 | 10811877011269 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 82 YR |