FDA Adverse Event Malfunction Summary report: N

MTOME ST HOLSTER/CABLES

MDR report key: 1162308 · Received September 8, 2008

Report

Report Number
3005075853-2008-01573
Event Type
Malfunction
Date Received
September 8, 2008
Report Date
July 1, 2008
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
KNW
PMA / PMN Number
K991980
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVAL SUMMARY: BASED ON ANALYSIS RESULTS, THIS COMPLAINT IS NOW DETERMINED TO BE A MDR MALFUNCTION. THE UNIT WAS RETURNED TO THE INTERNATIONAL SERVICE CENTER. BASED UPON THE INQUIRY INFO RECEIVED, VISUAL AND FUNCTIONAL EXAMINATION, THE UNIT WAS FOUND TO REQUIRE; DAMAGED UPPER AND LOWER CASE REPLACED, MISSING FASTENING MATERIAL EXCHANGED, ELECTRICAL, ROTATIONAL AND TRANSLATIONAL CABLES REPLACED. THE DEVICE HISTORY RECORD HAS BEEN REVIEWED VIA THE DATABASE. THE UNIT HAS PASSED ALL QA INSPECTIONS. COMPLAINT INFO IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED.

Description of Event or Problem · 1

IT WAS REPORTED BY THE SERVICE CENTER, THAT THE UNIT WAS SENT TO THEM FOR REPAIR, BECAUSE THE GREEN CABLE IS FAULTY. IT WAS NOT NOTED IF THE ISSUE OCCURRED DURING A PROCEDURE. NO PT CONSEQUENCE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MTOME ST HOLSTER/CABLES KNW ETHICON ENDO-SURGERY, LLC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 PROBE| CONTROL MODULE