FDA Adverse Event Malfunction Summary report: N

ZIMMER TRABECULAR METAL REVERSE SHOULDER SYSTEM GLENOSPHERE

MDR report key: 1162215 · Received September 3, 2008

Report

Report Number
1822565-2008-00576
Event Type
Malfunction
Date Received
September 3, 2008
Date of Event
July 23, 2008
Report Date
August 1, 2008
Manufacturer
ZIMMER, INC.
Product Code
HSD
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: THE GLENOSPHERE NOT SEATING PROPERLY ONTO BASEPLATE MAY BE CAUSED BY SOFT TISSUE, OR BONE TRAPPED AROUND THE BASE PLATE TAPER DURING GLENOSPHERE INSERTION, INSUFFICIENT BONE CLEARANCE AROUND THE BASE PLATE, INTERFERENCE WITH RETRACTORS, ETC. EACH SCENARIO COULD POTENTIALLY CAUSE THE GLENOSPHERE TO NOT COMPLETELY SEAT ON THE BASEPLATE. STRAIGHT-ON EXPOSURE OF THE GLENOID IS NECESSARY FOR BOTH PROPER REAMING AND COMPONENT INSERTION. THE USE OF BASE PLATE REAMER 2 (36MM OR 40MM) IS NEEDED TO REMOVE BONE AROUND THE BASE PLATE TO AVOID IMPINGEMENT WITH THE GLENOSPHERE. CAUSE CANNOT BE DEFINITIVELY DETERMINED. EVALUATION: NO PRODUCT WAS RETURNED. REVIEW OF THE DEVICE HISTORY RECORDS DID NOT FIND ANY DEVIATIONS OR ANOMALIES. IT IS NOT SUSPECTED THAT THE PRODUCT FAILED TO MEET SPECIFICATIONS. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY EVIDENCE OF PRODUCT CONTRIBUTION TO THE REPORTED PROBLEM. BASED ON THE INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. SHOULD ADDITIONAL SUBSTANTIVE INFORMATION BE RECEIVED, THE COMPLAINT WILL BE REOPENED. ZIMMER CONSIDERS THE INVESTIGATION CLOSED.

Description of Event or Problem · 1

IT IS REPORTED THAT THE DEVICE WAS IMPLANTED IN 2008. A POST-OP X-RAY REVEALED INCOMPLETE SEATING OF GLENOSPHERE ON BASEPLATE. IMPLANT WAS STABLE AND LOCKED ON AT SURGERY. STABLE ON POST-OPERATIVE SEQUENTIAL X-RAYS WITH NO CHANGE NOTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ZIMMER TRABECULAR METAL REVERSE SHOULDER SYSTEM GLENOSPHERE SHOULDER PROSTHESIS HSD ZIMMER, INC. NA 60834839

Patients

Seq Age Sex Outcome Treatment
1 89 YR