FDA Adverse Event Malfunction Summary report: N

FS DISPOSABLE INTERFACE

MDR report key: 11617184 · Received April 5, 2021

Report

Report Number
3006695864-2021-07442
Event Type
Malfunction
Date Received
April 5, 2021
Date of Event
February 12, 2021
Report Date
June 7, 2021
Manufacturer
AMO MANUFACTURING USA, LLC
Product Code
GEX
UDI-DI
15050474534688
PMA / PMN Number
K060372
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

SECTION D9: DEVICE AVAILABLE FOR EVALUATION: YES DATE RETURNED TO MANUFACTURER: 23 MAR, 2021 SECTION H3: DEVICE RETURNED TO MANUFACTURER: YES DEVICE EVALUATION: VISUAL INSPECTION WAS PERFORMED, AND NO OBVIOUS DAMAGE/DEFECT WAS OBSERVED. SUCTION TEST WAS PERFORMED, AND ALL RESULTS WERE FOUND WITHIN SPECIFICATIONS. THE REPORTED EVENT CANNOT BE CONFIRMED. A RECORD REVIEW WAS PERFORMED. A PRODUCT DEFICIENCY REVIEW WAS PERFORMED AND THERE IS NO PRODUCT DEFICIENCY IDENTIFIED. A DOCUMENT, LOT HISTORY, AND TRENDING WERE REVIEWED. THERE IS NOT A RECOGNIZABLE ADVERSE TREND. THE RISKS AND MITIGATIONS ASSOCIATED WITH THE COMPLAINT ISSUE ARE IDENTIFIED IN EXISTING RISK DOCUMENTS AND NO NEW RISKS WERE IDENTIFIED AS PART OF THIS INVESTIGATION. THE REVIEW OF THE DEVICE HISTORY RECORD (DHR) FOR THE PATIENT INTERFACE SHOWED THAT THERE WERE NO ISSUES OR NON-CONFORMITIES. THE DEVICE AND ITS COMPONENTS MET ALL SPECIFICATIONS PRIOR TO BEING RELEASED. MANUFACTURING HAS BEEN RULED OUT AS A POTENTIAL CAUSE FOR THE REPORTED ISSUE. BASED ON THE INVESTIGATION RESULTS, NO CORRECTIVE ACTION HAS BEEN ISSUED. CONCLUSION: BASED ON THE INFORMATION OBTAINED, PRODUCT MALFUNCTION AND PRODUCT DEFICIENCY CANNOT BE CONFIRMED. NO FURTHER INVESTIGATION IS REQUIRED ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED

Additional Manufacturer Narrative · 1

AGE OR DATE OF BIRTH, WEIGHT AND ETHNICITY: UNKNOWN/ NOT PROVIDED. THE PATIENT INTERFACE (PI) SUCTION RING MAY LOSE SUCTION DURING A PROCEDURE. LABEL COPY STATES CORNEAL FIXATION VACUUM LOSS CAN OCCUR. THERE ARE SEVERAL FACTORS THAT MAY CONTRIBUTE TO SUCTION ISSUES SUCH AS DOCTOR¿S TECHNIQUE IN APPLYING THE SUCTION RING TO THE CORNEA, DOCTOR¿S TECHNIQUE IN SQUEEZING THE PI CLIP TO SECURE THE SUCTION RING TO THE PI CONE AND PATIENT ANATOMY AFFECTING THE INTERFACE BETWEEN THE PATIENT¿S CORNEA AND THE SUCTION RING. IF THERE IS ANY FURTHER RELEVANT INFORMATION, A SUPPLEMENTAL MEDWATCH WILL BE FILED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED BY THE CUSTOMER THAT THEY EXPERIENCED SUCTION LOSS WITH THE PI DURING THE RASTER PATTERN OF LASER FIRING. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A NEW SUCTION RING ASSEMBLY. THERE WAS NO PATIENT INJURY OR SURGICAL INTERVENTION NEEDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
514219 FS DISPOSABLE INTERFACE POWERED LASER SURGICAL INSTRUMENT GEX AMO MANUFACTURING USA, LLC 590106AN 60276236 15050474534688

Patients

Seq Age Sex Outcome Treatment
1