COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP
Report
- Report Number
- 6000001-2007-00617
- Event Type
- Malfunction
- Date Received
- September 17, 2008
- Date of Event
- December 13, 2006
- Report Date
- December 13, 2006
- Manufacturer
- BAXTER HEALTHCARE PTE. LTD.
- Product Code
- FRN
- Removal / Correction Number
- 6000001-7/20/05-015-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER
Narratives
THIS REPORT IS BEING RESUBMITTED IN ACCORDANCE WITH INSTRUCTIONS FROM FDA TO ADDRESS A MANUFACTURER REPORT SEQUENCE NUMBER ISSUE THAT OCCURRED WHEN THIS MDR OR SUPPLEMENT WAS ORIGINALLY SUBMITTED TO THE FDA ON JAN 05 2007. EVALUATION SUMMARY: THE DEVICE EVALUATION WAS COMPLETED AND FAILURE CODE 703 WAS CONFIRMED (FOUND IN EVENT HISTORY) FOR CHANNEL C DUE TO A DEFECTIVE USER INTERFACE MODULE PRINTED CIRCUIT BOARD (UIM PCB). SERVICE INSTALLED A NEW UIM PCB, CHANNEL C AND TESTED THE DEVICE. A REVIEW OF THE COMPLAINT HISTORY REVEALED SIMILAR REPORTS HAVE BEEN RECEIVED FOR THIS PRODUCT AND THE REPORTED ISSUE. THIS ISSUE IS BEING INVESTIGATED UNDER CAPA.
THE DEVICE WAS RECEIVED FOR SERVICE. DURING SERVICE TESTING, FAILURE CODE 703 WAS FOUND IN TEH EVENT HISTORY. ACCORDING TO THE HOSPITAL REPRESENTATIVE THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE 3CX VOLUMETRIC INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE PTE. LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |