FDA Adverse Event Malfunction Summary report: N

ARTHROSCOPIC SHAVER, TORPEDO REGULAR LENGTH LIME GREEN

MDR report key: 11612999 · Received April 5, 2021

Report

Report Number
3014023545-2021-00004
Event Type
Malfunction
Date Received
April 5, 2021
Date of Event
January 8, 2021
Report Date
January 13, 2021
Manufacturer
RENOVO, INC.
Product Code
NLM
PMA / PMN Number
K172097
Removal / Correction Number
10054580-01082020-R-001
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
501

Narratives

Additional Manufacturer Narrative · 1

THE MANUFACTURER SHIPPED OUTER CORRUGATED BOXES TO THE CONTRACT STERILIZER THAT INCLUDE INNER BOXES THAT ARE ADDRESSED TO THE END CUSTOMER. THE COMPANY SHIPPED THE STERILIZER 6 OF THESE OUTER BOXES, BUT THE STERILIZER ONLY RECEIVED 5 OUT OF THE 6 BOXES. THE COMPANY WAITED SEVERAL DAYS TO SEE IF THE 6TH BOX WOULD ARRIVE AT THE STERILIZER, BUT IT NEVER DID. AT THAT POINT, THE COMPANY INITIATED A CLAIM WITH (B)(6). IN LOOKING AT THE (B)(6) RECORDS, THE OUTER PACKAGE WAS RECEIVED AT A (B)(6) FACILITY ON 12/21. ON THE 22ND, THE INNER BOXES WERE PICKED UP FROM THAT (B)(6) LOCATION AND SHIPPED TO THE CUSTOMERS. IT IS UNKNOWN WHY THE OUTER BOX (ADDRESSED TO THE STERILIZER) WAS OPENED BY (B)(6) TO EXPOSE THE 3 INNER BOXES. AS SOON AS THE COMPANY WAS AWARE THAT NON-STERILE COMPONENTS HAD BEEN DELIVERED TO THE END USERS THEY REACHED OUT TO INSTRUCT THE FACILITIES TO IMMEDIATELY QUARANTINE THE DEVICES AND SHIP THEM BACK TO THE MANUFACTURER AS SOON AS POSSIBLE. TWO OF THE THREE SITES WHO HAD RECEIVED THE NON-STERILE PRODUCT HAD USED AT LEAST ONE OF THE DEVICES. THE FACILITY WAS ABLE TO NARROW IT DOWN TO SEVERAL PATIENTS WHO COULD HAVE BEEN EXPOSED BUT NOT THE EXACT SINGLE PATIENT.

Description of Event or Problem · 1

A NON-STERILE SHAVER WAS USED ON A PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
514328 ARTHROSCOPIC SHAVER, TORPEDO REGULAR LENGTH LIME GREEN NLM RENOVO, INC. AS-AR-8400TD L-008936- 16275

Patients

Seq Age Sex Outcome Treatment
1 60 YR