FDA Adverse Event Injury Summary report: N

STARCLOSE VASCULAR CLOSURE SYSTEM

MDR report key: 1161286 · Received September 15, 2008

Report

Report Number
2953144-2008-01484
Event Type
Injury
Date Received
September 15, 2008
Date of Event
August 19, 2008
Report Date
August 20, 2008
Manufacturer
ABBOTT VASCULAR REDWOOD CITY
Product Code
MGB
PMA / PMN Number
P050007
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER REPORTED THE DEVICE WAS DISCARDED. THE LOT NUMBER WAS NOT IDENTIFIED; THEREFORE, A DEVICE HISTORY RECORD REVIEW COULD NOT BE PERFORMED. A POWER POINT SLIDE SET OF 5 ANGIOGRAM IMAGES WAS PROVIDED AND WERE REVIEWED BY ABBOTT VASCULAR IN-HOUSE MEDICAL MONITOR. BASED ON THE FINDINGS, THE IMAGES SHOW DIMINISHED FLOW AT THE LEVEL OF THE CLIP; HOWEVER, THE CAUSE CANNOT BE SEEN FROM THE IMAGES PROVIDED. THESE IMAGES DO NOT RULE OUT NOR CONFIRM THAT THE CLIP WAS ATTACHED TO THE ANTERIOR AND POSTERIOR ARTERY WALLS AND THE ROOT CAUSE IS UNDETERMINED.

Description of Event or Problem · 1

DEVICE ISSUE: CLIP MISLOCATION. TIME OF DEVICE ISSUE: DURING VESSEL CLOSURE. SYMPTOMS/AE: FAILURE TO ACHIEVE HEMOSTASIS, PSEUDOANEURYSM, ISCHEMIA, DISSECTION, ENDARTERECTOMY AND THROMBECTOMY. IT WAS REPORTED THAT A PHYSICIAN TRAINED IN THE USE OF THE STARCLOSE DEVICE ATTEMPTED ARTERIOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL ARTERY (CFA) AFTER AN INTERVENTIONAL PROCEDURE. REPORTEDLY, ANGIOGRAPHY SHOWED A TIGHT "STENOSIS" IN THE DISTAL RIGHT SUPERFICIAL FEMORAL ARTERY (SFA) UNDERLYING THE SITE OF THE STARCLOSE WITH PROBABLE DISSECTION. ADDITIONALLY, AN OCCLUDED MID-DISTAL POPLITEAL ARTERY WITH NO TIBIAL RUN OFF WAS FOUND CONSISTENT WITH ACUTE THROMBUS. THE PT WAS TAKEN TO SURGERY. REPORTEDLY, THE CLIP WAS FOUND ATTACHED TO THE ANTERIOR AND POSTERIOR WALL. THE PT EXHIBITED A COLD, PALE, PULSELESS RIGHT FOOT (ABSENT DORSALIS PEDIS AND POPLITEAL ARTERY PULSES.) A PSEUDOANEURYSM WAS FOUND AT THE RIGHT COMMON FEMORAL ARTERY. THROMBECTOMY WAS PERFORMED FOLLOWED BY A COMMON FEMORAL ENDARTERECTOMY AND VEIN GRAFT REPAIR. THERE WERE NO REPORTED ADVERSE PT SEQUELAE. NO ADDITIONAL INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STARCLOSE VASCULAR CLOSURE SYSTEM MGB ABBOTT VASCULAR REDWOOD CITY NA UNK

Patients

Seq Age Sex Outcome Treatment
1 68 YR Required Intervention