FDA Adverse Event Malfunction Summary report: N

BUDDY LITE WITH AC, BATTERY HOUSING AND HEATER UNIT, N. AMERICA

MDR report key: 11612730 · Received April 5, 2021

Report

Report Number
11612730
Event Type
Malfunction
Date Received
April 5, 2021
Date of Event
March 30, 2021
Report Date
April 1, 2021
Manufacturer
BELMONT INSTRUMENT CORPORATION
Product Code
BSB
UDI-DI
10896128002449
Product Problem
Yes
Report Source
User Facility report
Reporter Location
FL
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

THE AC ADAPTER OF THE BELMONT BUDDY LITE FLUID WARMER BECAME WET WITH BLOOD. WHILE TRYING TO CLEAN THE DEVICE, THE USER USED CAVICIDE AND SPRAYED THE AC ADAPTER WITHOUT UNPLUGGING IT FROM THE ELECTRICAL OUTLET, CAUSING A SHORT CIRCUIT ON THE POWER SUPPLY THAT GENERATED SPARKS AND SMOKE IN THE OPERATING ROOM (OR). THE OPERATING ROOM PROCEDURE WAS ALREADY COMPLETED AT THE TIME OF THE INCIDENT, NO PATIENT OR USER WERE HARMED BY THIS SITUATION. THE AC ADAPTER AND BELMONT BUDDY LITE WERE REMOVED FROM SERVICE. OPERATING ROOM LEADERSHIP CONDUCTED A PROCEDURE REVIEW WITH OPERATING ROOM PERSONNEL TO REVIEW PROPER CLEANING PROCEDURES OF EQUIPMENT THAT ARE CONTAMINATED WITH BIOLOGICAL FLUIDS.