FDA Adverse Event
Malfunction
Summary report: N
BUDDY LITE WITH AC, BATTERY HOUSING AND HEATER UNIT, N. AMERICA
MDR report key: 11612730
·
Received April 5, 2021
Report
- Report Number
- 11612730
- Event Type
- Malfunction
- Date Received
- April 5, 2021
- Date of Event
- March 30, 2021
- Report Date
- April 1, 2021
- Manufacturer
- BELMONT INSTRUMENT CORPORATION
- Product Code
- BSB
- UDI-DI
- 10896128002449
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- FL
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Description of Event or Problem · 1
THE AC ADAPTER OF THE BELMONT BUDDY LITE FLUID WARMER BECAME WET WITH BLOOD. WHILE TRYING TO CLEAN THE DEVICE, THE USER USED CAVICIDE AND SPRAYED THE AC ADAPTER WITHOUT UNPLUGGING IT FROM THE ELECTRICAL OUTLET, CAUSING A SHORT CIRCUIT ON THE POWER SUPPLY THAT GENERATED SPARKS AND SMOKE IN THE OPERATING ROOM (OR). THE OPERATING ROOM PROCEDURE WAS ALREADY COMPLETED AT THE TIME OF THE INCIDENT, NO PATIENT OR USER WERE HARMED BY THIS SITUATION. THE AC ADAPTER AND BELMONT BUDDY LITE WERE REMOVED FROM SERVICE. OPERATING ROOM LEADERSHIP CONDUCTED A PROCEDURE REVIEW WITH OPERATING ROOM PERSONNEL TO REVIEW PROPER CLEANING PROCEDURES OF EQUIPMENT THAT ARE CONTAMINATED WITH BIOLOGICAL FLUIDS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 513298 | BUDDY LITE WITH AC, BATTERY HOUSING AND HEATER UNIT, N. AMERICA | WARMER, BLOOD, NON-ELECTROMAGNETIC RADIATION | BSB | BELMONT INSTRUMENT CORPORATION | 905-00022 | 10896128002449 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |